NKII POROUS COATED STEMMED TIBIAL BASEPLATE
Report
- Report Number
- 1822565-2011-01491
- Event Type
- Injury
- Date Received
- June 23, 2011
- Date of Event
- April 5, 2011
- Report Date
- May 27, 2011
- Manufacturer
- ZIMMER INC.
- Product Code
- JWH
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- PHYSICIAN
Narratives
EVAL SUMMARY: NO PRODUCT OR X-RAYS WERE RETURNED FOR REVIEW. WHILE A CAUSE FOR THIS SPECIFIC CLINICAL EVENT CANNOT BE ASCERTAINED FROM THE INFO PROVIDED, THE COMMON CLINICAL PRESENTATION SUGGESTS THAT THE EVENT IS RELATED TO THE ISSUES FOR WHICH ZIMMER IMPLEMENTED A CORRECTION ACTION FOR IN 2007. THIS CORRECTIVE ACTION INVOLVED ENHANCING THE LABELING FOR THE NATURAL KNEE II KNEE SYSTEM. REFERENCE MDR 1822565-2006-00284 FOR ADDITIONAL INFO REGARDING THE CORRECTIVE ACTION TAKEN. EVAL: NO PRODUCT WAS RETURNED. REVIEW OF THE DEVICE HISTORY RECORDS DID NOT FIND ANY DEVIATIONS OR ANOMALIES. IT IS NOT SUSPECTED THAT THE PRODUCT FAILED TO MEET SPECIFICATIONS. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY EVIDENCE OF PRODUCT CONTRIBUTION TO THE REPORTED PROBLEM. SHOULD ADDITIONAL SUBSTANTIVE INFO BE RECEIVED, THE COMPLAINT WILL BE REOPENED.
IT IS REPORTED THAT THE PT WAS REVISED DUE TO OSTEOLYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NKII POROUS COATED STEMMED TIBIAL BASEPLATE | KNEE PROSTHESIS | JWH | ZIMMER INC. | 1537760 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | CATALOG #620008811, LOT #1457418| NATURAL-KNEE II CONGRUENT TIBIAL INSERT: |