FDA Adverse Event Injury Summary report: N

GMK REVISION TOTAL KNEE SYSTEM

MDR report key: 21479207 · Received February 27, 2025

Report

Report Number
3005180920-2025-00126
Event Type
Injury
Date Received
February 27, 2025
Date of Event
February 1, 2025
Report Date
February 27, 2025
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JWH
PMA / PMN Number
K103170
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 07 FEBRUARY 2025. LOT 2404816: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 24-APRIL-2024. EXPIRATION DATE: 2029-03-21. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. INFECTION IS A KNOWN POSSIBLE COMPLICATION OF ANY SURGERY. ALTHOUGH NO ROOT CAUSE CAN BE ESTABLISHED, THERE IS NO INDICATION THAT ANY ISSUE WITH THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AND THE DOCUMENT REVIEW DOES NOT INDICATE ANY POTENTIAL MANUFACTURING RELATED CAUSE.

Description of Event or Problem · 0

AT ABOUT 4 MONTHS AFTER THE PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1123696 GMK REVISION TOTAL KNEE SYSTEM TIBIAL INSERT SC FIXED 10MM, SIZE 1 JWH MEDACTA INTERNATIONAL SA 02.07.0110SCF 2404816

Patients

Seq Age Sex Outcome Treatment
1 72 YR Female Required Intervention