ONDASPACE EXTENSION ARM
Report
- Report Number
- 2031963-2011-00045
- Event Type
- Malfunction
- Date Received
- June 10, 2011
- Date of Event
- May 16, 2011
- Report Date
- June 10, 2011
- Manufacturer
- STRYKER COMMUNICATIONS
- Product Code
- BRY
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THERE IS NO EXPIRATION DATE FOR THIS PRODUCT. THE SERIAL NUMBER IS UNKNOWN SO THE DEVICE MANUFACTURE DATE IS UNKNOWN. DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER AT THE TIME OF THIS REPORT. EVALUATION SUMMARY: IT WAS REPORTED THAT A 182MM REDUCING RING FROM CEILING PLATE FOR THE DUAL FLAT PANEL FELL OFF. IT WAS FURTHER STATED THAT THE RING HIT A HOSPITAL EMPLOYEE. UPON FURTHER INVESTIGATION IT WAS FOUND THAT THE RING THAT FELL WAS A BLANK REDUCER RING. STRYKER PROVIDES SEVERAL CONFIGURATIONS FOR MOUNTING THE SUSPENSIONS AND ONE OF THE CONFIGURATIONS IS A TANDEM MOUNTING PLATE WHICH ALLOWS TWO SUSPENSIONS OR A SUSPENSION AND BOOM TO BE MOUNTED RIGHT NEXT TO EACH OTHER. IN THIS OPERATING ROOM, THERE IS NOTHING MOUNTED NEXT TO THE OTHER SUSPENSION. ALTHOUGH THERE IS NO OTHER DEVICE MOUNTED NEXT TO THE SUSPENSION IN A TANDEM CONFIGURATION, A TANDEM CEILING COVER IS USED TO CONCEAL THE HOLE IN THE CEILING. THIS COVER IS OFTEN REFERRED TO AS A REDUCER RING OR IN THIS CASE, A BLANK REDUCER RING. IT APPEARS AS IF THE CAUSE OF THE REDUCER RING TO FALL WAS A SPRING ARM TENSION ISSUE CAUSING THE SPRING ARM OF THE LIGHTS WITHIN THE ROOM TO CONTINUE TO DRIFT UPWARDS AND IMPACT THE CEILING COVER. THIS LIKELY CAUSED THE CEILING COVER TO VIBRATE WHICH EVENTUALLY CAUSED THE REDUCER RING COVER TO COME LOOSE AND FALL. THE CEILING COVER BLANK REDUCER RING IS ATTACHED TO THE CEILING COVER VIA SEVERAL SMALL METAL FASTENING TABS AND PRESS FITS INTO ITS LOCATION. THE REDUCER RING IS A METAL COVER THAT ALLEGEDLY IMPACTED A NURSE WHEN IT FELL DURING A CASE. WHILE THERE WAS A REPORT OF A STAFF MEMBER BEING INJURED, IT IS UNKNOWN AT THIS TIME WHO THE STAFF MEMBER WAS OR WHETHER THIS CAUSED A SERIOUS OR SEVERE INJURY. THIS NOT A SINGLE USE DEVICE.
STRYKER REFERENCE#: (B)(4). (B)(6) UNIVERSITY: (B)(6) CALLED IN 182MM REDUCING RING FROM CEILING PLATE FOR THE DUAL FLAT PANEL FELL OFF. RING HIT HOSPITAL EMPLOYEE. ACCOUNT CLOSING ROOM UNTIL STRYKER CAN VERIFY STRUCTURE IS SAFE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONDASPACE EXTENSION ARM | BRY | STRYKER COMMUNICATIONS | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |