FDA Adverse Event Injury Summary report: N

ONE PEG PATELLA 32MM

MDR report key: 21292988 · Received February 3, 2025

Report

Report Number
1038671-2025-00614
Event Type
Injury
Date Received
February 3, 2025
Date of Event
April 4, 2023
Report Date
February 26, 2025
Manufacturer
EXACTECH, INC.
Product Code
JWH
UDI-DI
10885862039651
PMA / PMN Number
K932690
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE REASON FOR THE REVISION REPORTED CANNOT BE CONFIRMED FROM THE INFORMATION PROVIDED BUT MAY BE THE RESULT OF PROSTHESIS WEAR AND PATELLAR LOOSENING OR DUE TO INCLUSION OF THE POLYETHYLENE IN THE PACKAGING RECALL. POTENTIAL CONTRIBUTIONS OF USER AND PATIENT-RELATED CONSIDERATIONS TO THE EVENT COULD NOT BE ASSESSED AS THE DEVICES WERE NOT AVAILABLE FOR EVALUATION AND IMAGES AND RADIOGRAPHS WERE NOT PROVIDED. H6: CORRECTED HEALTH EFFECT CLINICAL CODES.

Additional Manufacturer Narrative · 0

D10: (B)(6), 201-78-15 - HOLDING PIN MINI SHARP POINT 4 PK. (B)(6), 02-010-03-0340 - LOGIC CR FEMORAL CEM, RIGHT, SZ 4 . (B)(6), 02-012-45-4040 - LGC TIBIAL FIT TRAY CEM SZ 4F / 4T. THE PRODUCT ASSOCIATED WITH THE REPORTED EVENT IS WITHIN THE SCOPE OF RECALL 1038671-04/18/2024-002-R; HOWEVER, THERE IS INSUFFICIENT INFORMATION TO EVALUATE WHETHER THE SUBJECT ISSUE OF THE RECALL WAS A CAUSE OR CONTRIBUTOR TO THE REPORTED EVENT. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORT: 1038671-2024-01465. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS; HOWEVER, AN INVESTIGATION OF THE REPORTED EVENT IS IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETED, A SUPPLEMENTAL MEDWATCH 3500A WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT APPROXIMATELY 119 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, OSTEOLYSIS, PATELLAR COMPONENT LOOSENING, PAIN AND SWELLING. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1157279 ONE PEG PATELLA 32MM PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL JWH EXACTECH, INC. 10885862039651

Patients

Seq Age Sex Outcome Treatment
1 76 YR Male Required Intervention| H SEE H11