THREE PEG PATELLA 32MM
Report
- Report Number
- 1038671-2025-00348
- Event Type
- Injury
- Date Received
- January 22, 2025
- Date of Event
- November 13, 2013
- Report Date
- January 22, 2025
- Manufacturer
- EXACTECH, INC.
- Product Code
- JWH
- UDI-DI
- 10885862039590
- PMA / PMN Number
- K932690
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
1038671-2024-01486 D10: CONCOMITANT DEVICES (B)(6), 02-010-01-0325 - LOGIC FEMORAL PS CEM RIGHT SZ 2.5, (B)(6), 02-012-45-2525 - LGC TIBIAL FIT TRAY CEM SZ 2.5F / 2.5T. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORT: 1038671-2024-01486. THE REASON FOR THE REVISION REPORTED CANNOT BE CONFIRMED FROM THE INFORMATION PROVIDED BUT MAY BE THE RESULT OF DISLOCATION, PROSTHESIS WEAR, AND INSTABILITY. OR DUE TO INCLUSION OF THE POLYETHYLENE IN THE PACKAGING RECALL. POTENTIAL CONTRIBUTIONS OF USER AND PATIENT-RELATED CONSIDERATIONS TO THE EVENT COULD NOT BE ASSESSED AS THE DEVICES WERE NOT AVAILABLE FOR EVALUATION AND IMAGES AND RADIOGRAPHS WERE NOT PROVIDED. IF ANY FURTHER INFORMATION IS OBTAINED THAT WOULD CHANGE OR ALTER ANY INFORMATION PROVIDED, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.
IT WAS REPORTED THAT APPROXIMATELY 6 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PATELLAR SUBLUXATION, INSTABILITY, SCARRING, MILD INTERNAL ROTATION OF TIBIAL COMPONENT. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 683424 | THREE PEG PATELLA 32MM | PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL | JWH | EXACTECH, INC. | 10885862039590 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR | Unknown | Required Intervention| H | SEE H11 |