FDA Adverse Event Injury Summary report: N

LOGIC CR TIB INSERT SLOPE ++, SZ 2.5, 9 MM

MDR report key: 20966150 · Received December 18, 2024

Report

Report Number
1038671-2024-04921
Event Type
Injury
Date Received
December 18, 2024
Date of Event
February 12, 2022
Report Date
March 10, 2025
Manufacturer
EXACTECH, INC.
Product Code
JWH
PMA / PMN Number
K111400
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

D10. 4863565, 02-010-03-0325 - LOGIC CR FEMORAL CEM, RIGHT, SZ 2.5, 5127707, 02-012-45-2525 - LGC TIBIAL FIT TRAY CEM SZ 2.5F / 2.5T, (B)(6), 02-012-49-2509 - LOGIC CR TIB INSERT SLOPE ++, SZ 2.5, 9 MM, 5083388, 200-02-32 - THREE PEG PATELLA 32MM. THE REPORTED EVENT WAS UNABLE TO BE CONFIRMED DUE TO LIMITED INFORMATION RECEIVED FROM THE CUSTOMER. NO DEVICE WAS RETURNED FOR EVALUATION; FURTHER, PHOTOGRAPHS AND/OR RADIOGRAPH IMAGES WERE NOT PROVIDED FOR REVIEW. OPERATIVE NOTES AND/OR MEDICAL RECORDS WERE NOT PROVIDED FOR REVIEW OF USAGE/ TECHNIQUE. A DEFINITIVE ROOT CAUSE WAS UNABLE TO BE DETERMINED AS THE NECESSARY INFORMATION TO ADEQUATELY INVESTIGATE THE REPORTED EVENT WAS NOT PROVIDED. IF ANY FURTHER INFORMATION IS OBTAINED THAT WOULD CHANGE OR ALTER ANY INFORMATION PROVIDED, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL AND/OR CORRECTED INFORMATION. THE REASON FOR THE REVISION REPORTED CANNOT BE CONFIRMED FROM THE INFORMATION PROVIDED BUT MAY BE THE RESULT OF PROSTHESIS WEAR OR DUE TO INCLUSION OF THE POLYETHYLENE IN THE PACKAGING RECALL. ADDITIONALLY, A CONTRIBUTING FACTOR TO THE REPORTED WEAR MAY HAVE BEEN THE RESULT OF BEING PACKAGED IN A NON-CONFORMING VACUUM BAG FOR MORE THAN FIVE YEARS. POTENTIAL CONTRIBUTIONS OF USER AND PATIENT-RELATED CONSIDERATIONS TO THE EVENT COULD NOT BE ASSESSED AS THE DEVICES WERE NOT AVAILABLE FOR EVALUATION AND IMAGES, RADIOGRAPHS, AND RELEVANT CLINICAL INFORMATION WERE NOT PROVIDED. IF ANY FURTHER INFORMATION IS OBTAINED THAT WOULD CHANGE OR ALTER ANY INFORMATION PROVIDED, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Description of Event or Problem · 0

IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 50 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2082429 LOGIC CR TIB INSERT SLOPE ++, SZ 2.5, 9 MM PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL JWH EXACTECH, INC.

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention| H SEE H11