LOGIC CR TIB INSERT SLOPE ++, SZ 2.5, 9 MM
Report
- Report Number
- 1038671-2024-04921
- Event Type
- Injury
- Date Received
- December 18, 2024
- Date of Event
- February 12, 2022
- Report Date
- March 10, 2025
- Manufacturer
- EXACTECH, INC.
- Product Code
- JWH
- PMA / PMN Number
- K111400
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
D10. 4863565, 02-010-03-0325 - LOGIC CR FEMORAL CEM, RIGHT, SZ 2.5, 5127707, 02-012-45-2525 - LGC TIBIAL FIT TRAY CEM SZ 2.5F / 2.5T, (B)(6), 02-012-49-2509 - LOGIC CR TIB INSERT SLOPE ++, SZ 2.5, 9 MM, 5083388, 200-02-32 - THREE PEG PATELLA 32MM. THE REPORTED EVENT WAS UNABLE TO BE CONFIRMED DUE TO LIMITED INFORMATION RECEIVED FROM THE CUSTOMER. NO DEVICE WAS RETURNED FOR EVALUATION; FURTHER, PHOTOGRAPHS AND/OR RADIOGRAPH IMAGES WERE NOT PROVIDED FOR REVIEW. OPERATIVE NOTES AND/OR MEDICAL RECORDS WERE NOT PROVIDED FOR REVIEW OF USAGE/ TECHNIQUE. A DEFINITIVE ROOT CAUSE WAS UNABLE TO BE DETERMINED AS THE NECESSARY INFORMATION TO ADEQUATELY INVESTIGATE THE REPORTED EVENT WAS NOT PROVIDED. IF ANY FURTHER INFORMATION IS OBTAINED THAT WOULD CHANGE OR ALTER ANY INFORMATION PROVIDED, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL AND/OR CORRECTED INFORMATION. THE REASON FOR THE REVISION REPORTED CANNOT BE CONFIRMED FROM THE INFORMATION PROVIDED BUT MAY BE THE RESULT OF PROSTHESIS WEAR OR DUE TO INCLUSION OF THE POLYETHYLENE IN THE PACKAGING RECALL. ADDITIONALLY, A CONTRIBUTING FACTOR TO THE REPORTED WEAR MAY HAVE BEEN THE RESULT OF BEING PACKAGED IN A NON-CONFORMING VACUUM BAG FOR MORE THAN FIVE YEARS. POTENTIAL CONTRIBUTIONS OF USER AND PATIENT-RELATED CONSIDERATIONS TO THE EVENT COULD NOT BE ASSESSED AS THE DEVICES WERE NOT AVAILABLE FOR EVALUATION AND IMAGES, RADIOGRAPHS, AND RELEVANT CLINICAL INFORMATION WERE NOT PROVIDED. IF ANY FURTHER INFORMATION IS OBTAINED THAT WOULD CHANGE OR ALTER ANY INFORMATION PROVIDED, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.
IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 50 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2082429 | LOGIC CR TIB INSERT SLOPE ++, SZ 2.5, 9 MM | PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL | JWH | EXACTECH, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Required Intervention| H | SEE H11 |