FDA Adverse Event Other Summary report: N

LAMICUS SEPARATOR SYSTEM

MDR report key: 2093992 · Received April 26, 2011

Report

Report Number
3004548776-2011-00001
Event Type
Other
Date Received
April 26, 2011
Date of Event
April 1, 2011
Report Date
April 1, 2011
Manufacturer
FENWAL, INC.
Product Code
GKT
PMA / PMN Number
BK960005
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

PLEASE NOTE THAT THE OPERATORS MANUAL INDICATES A WARNING AND A CAUTION REGARDING THE PRIMING OF THE KIT. WARNING: OPENING THE RETURN LINE ROLLER CLAMP AFTER VENIPUNCTURE AND BEFORE THE BLOOD SAMPLING POUCH IS FILLED MAY RESULT IN AN AIR EMBOLISM. CAUTION: THE KIT SHOULD BE ALMOST COMPLETELY FULL OF FLUID (SALINE AND ACD) AT THE END OF PRIME. THE CONTAINERS ON THE FRONT SCALE HOOKS SHOULD HAVE LITTLE OR NO FLUID IN THEM, EXCEPT FOR THE CONTAINER ON THE FRONT LEFT SCALE HOOK ON A DOUBLE NEEDLE OR MNC KIT, OR THE CONTAINER ON THE FRONT RIGHT SCALE HOOK ON A SINGLE NEEDLE KIT. SOME AIR SHOULD BE EXPECTED NEAR THE FISTULA(S). IF THERE ARE LARGE AMOUNTS OF AIR IN THE KIT IMMEDIATELY FOLLOWING PRIME, DO NOT BEGIN COLLECTION. REMOVE THE KIT AND TRY INSTALLING AND PRIMING ANOTHER KIT. (B)(4).

Description of Event or Problem · 1

A CUSTOMER REPORTED WHILE SETTING UP A DOUBLE NEEDLE AMICUS PROCEDURE THE RETURN LINE WAS NOT PRIMED BY THE OPERATOR PRIOR TO VENIPUNCTURE AND APPROXIMATELY 1.2 ML OF AIR WAS INFUSED. DONOR DID NOT EXPERIENCE ANY SYMPTOMS OF AIR EMBOLISM. THE PROCEDURE WAS DISCONTINUED IMMEDIATELY. THE FACILITIES MEDICAL DIRECTOR WAS INFORMED OF THE SITUATION AND DONOR FOLLOW UP WAS RECOMMENDED. DONOR CONDITION WAS STABLE AND LEFT ON OWN ACCORD. THE CENTER FOLLOWED UP WITH THE DONOR AT A LATER DATE AND THERE WERE NO ADVERSE EFFECTS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LAMICUS SEPARATOR SYSTEM AUTOMATED BLOOD CELL SEPARATOR GKT FENWAL, INC. NA FA11A17173

Patients

Seq Age Sex Outcome Treatment
1 41 YR Other LOT FA11A17173| AMICUS APHERESIS KIT DOUBLE NEEDLE: (B)(4)