UNKNOWN ZIMMER KNEE
Report
- Report Number
- 1822565-2011-01141
- Event Type
- Injury
- Date Received
- May 10, 2011
- Date of Event
- December 22, 2009
- Report Date
- April 12, 2011
- Manufacturer
- ZIMMER, INC.
- Product Code
- JWH
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- PHYSICIAN
Narratives
INFORMATION WAS RECEIVED FROM A HEALTH PROFESSIONAL WHO IS NOT REQUIRED TO COMPLETE FORM 3500A. EVALUATION SUMMARY: IT HAS BEEN REPORTED THAT THE PATIENT WAS REVISED APPROXIMATELY 1 YEAR POST OP. NO DEVICES OR PHOTOS WERE RECEIVED; THEREFORE, THE CONDITION OF THE COMPONENTS IS UNKNOWN. THE SUPPLIED OPERATIVE NOTES INDICATE SOME LAXITY TO VARUS STRESS AND TIBIAL TRANSLATION. HOWEVER, X-RAYS WERE NOT PROVIDED; IT IS UNKNOWN WHETHER THE COMPONENTS WERE IMPLANTED WITH THE CORRECT FIT AND ORIENTATION AS PER THE SURGICAL TECHNIQUE. PATIENT FACTORS THAT MAY AFFECT THE PERFORMANCE OF THE COMPONENTS SUCH AS BONE QUALITY, HEIGHT/WEIGHT, ACTIVITY LEVEL, TYPE OF ACTIVITY (LOW IMPACT VS. HIGH IMPACT) ARE UNKNOWN. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED WITH THE INFORMATION PROVIDED. EVALUATION: REVIEW OF THE DEVICE HISTORY RECORDS WAS NOT POSSIBLE AS THE PRODUCT AND/OR LOT NUMBERS REQUIRED FOR RETRIEVAL WERE UNAVAILABLE. IT IS NOT SUSPECTED THAT THE PRODUCT FAILED TO MEET SPECIFICATIONS. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY EVIDENCE OF PRODUCT CONTRIBUTION TO THE REPORTED PROBLEM. BASED ON THE AVAILABLE INFORMATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. SHOULD ADDITIONAL SUBSTANTIVE INFORMATION BE RECEIVED, THE COMPLAINT WILL BE REOPENED. ZIMMER, INC CONSIDERS THE INVESTIGATION CLOSED.
IT IS REPORTED THAT THE PATIENT WAS REVISED FOR INSTABILITY, PAIN, AND RECURRENT EFFUSIONS APPROXIMATELY ONE YEAR AFTER THE DEVICES WERE IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNKNOWN ZIMMER KNEE | KNEE PROSTHESIS | JWH | ZIMMER, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 51 YR | Required Intervention |