FDA Adverse Event Malfunction Summary report: N

DIRECT PRESSURE MONITOR

MDR report key: 207912 · Received January 25, 1999

Report

Report Number
2023988-1999-00003
Event Type
Malfunction
Date Received
January 25, 1999
Date of Event
January 11, 1999
Report Date
January 25, 1999
Manufacturer
NEUROCARE GROUP, CAMINO
Product Code
GWM
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

INTRACRANIAL PRESSURE VALUE WAS ABOUT -50 ON NOVEMBER 13, WHEN THE 10 DAYS PASSED SINCE WE STARTED TO USE 110-4HMT. WHEN WE REPLACED THE MONITOR WITH ANOTHER ONE (MPH-1), WE COULD SEE THE VALUE (ABOUT 25) SAME AS BEFORE. THEREFORE, THERE MAY BE SOMETHING WRONG IN THE MAIN UNIT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DIRECT PRESSURE MONITOR ICP MONITOR GWM NEUROCARE GROUP, CAMINO V420-9 NA

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Required Intervention