FDA Adverse Event Injury Summary report: N

SEBRA SEALER 2380 AND SEBRA SEALER HEAD 1105

MDR report key: 2030904 · Received March 21, 2011

Report

Report Number
1219343-2011-00003
Event Type
Injury
Date Received
March 21, 2011
Date of Event
February 17, 2011
Report Date
March 14, 2011
Manufacturer
HAEMONETICS CORP.
Product Code
GKT
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

ADD'L INFO: DEVICE MANUFACTURE DATE: (B)(4) 1994. THE EQUIPMENT WAS RETURNED TO HAEMONETICS FOR EVAL ON (B)(4) 2011. THE SEALER POWER SUPPLY, (B)(4) WAS EVALUATED. THE SEALER POWER SUPPLY OPERATED CORRECTLY AND NO DEFECT WAS FOUND. THE HAND HELD SEALING HEAD, (B)(4) WAS FOUND TO BE OUT OF SPECIFICATION FOR RESISTANCE AND PHASE ANGLE. THE JAWS WERE FOUND TO BE DIRTY AND THE EQUIPMENT REQUIRED MAINTENANCE. THE EQUIPMENT DID NOT OPERATE CORRECTLY, AS AN ELECTRICAL ARC WAS GENERATED DURING OPERATION. THE JAWS, JAW HOLDER, AND MAGNETIC SWITCH WERE REPLACED. THE SEALING HEAD WAS DAMAGED DUE TO ARCING, CAUSED BY CONTAMINATION OF THE SEALING REGION. AFTER SERVICE, THE EQUIPMENT OPERATED CORRECTLY. CLEANING INSTRUCTIONS ARE COVERED IN THE MANUAL AS WELL AS TROUBLESHOOTING INSTRUCTIONS FOR DAMAGED JAW. ALSO, THE MANUAL CONTAINS A WARNING ASSOCIATED WITH ARCING, WHICH MAY LEAD TO BURNS IF THE EQUIPMENT IS NOT CLEANED AND/OR MAINTENANCE PERFORMED IN ACCORDANCE WITH (B)(4), MANUAL REVISION (B)(4). THIS INCIDENT IS DUE TO FAILURE TO MAINTAIN THE EQUIPMENT.

Description of Event or Problem · 1

THE NURSE, WHILE SEALING A TUBE FILLED WITH SALINE SOLUTION, EXPERIENCED PAIN WHILE OBSERVING SPARKS BETWEEN HER FOREFINGER AND THE SEALING APPARATUS. A MEDICAL CERTIFICATE FROM (B)(6) 2011 CONFIRMS A BURN WITH PAIN DIAGNOSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SEBRA SEALER 2380 AND SEBRA SEALER HEAD 1105 TUBING SEALER AND SEALER HEAD GKT HAEMONETICS CORP. 2380/1105

Patients

Seq Age Sex Outcome Treatment
1 NA