NEXGEN MIS STEMMED TIBIAL COMPONENT
Report
- Report Number
- 1822565-2010-01344
- Event Type
- Injury
- Date Received
- December 14, 2010
- Date of Event
- November 15, 2010
- Report Date
- November 15, 2010
- Manufacturer
- ZIMMER, INC.
- Product Code
- JWH
- Removal / Correction Number
- 1822565-04/19/2010-001
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
INFO WAS RECEIVED FROM A CUSTOMER WHO IS NOT REQUIRED TO COMPLETE FORM 3500A. EVAL SUMMARY: A CAUSE FOR THIS SPECIFIC CLINICAL EVENT CANNOT BE ASCERTAINED FROM THE INFO PROVIDED. WHILE A CAUSE FOR THIS SPECIFIC CLINICAL EVENT CANNOT BE ASCERTAINED FROM THE INFO PROVIDED, THE COMMON CLINICAL PRESENTATION AND THE DATE OF THE ORIGINAL IMPLANTATION SUGGEST THAT THE REVISION SURGERY MAY BE RELATED TO THE ISSUES FOR WHICH ZIMMER IMPLEMENTED A CORRECTION ACTION IN (B)(6) 2010. EVAL: REVIEW OF THE DEVICE HISTORY RECORDS DID NOT FIND ANY DEVIATIONS OR ANOMALIES. IT IS NOT SUSPECTED THAT THE PRODUCT FAILED TO MEET SPECIFICATIONS. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY EVIDENCE OF PRODUCT CONTRIBUTION TO THE REPORTED PROBLEM. A FIELD ACTION WAS CONDUCTED ON (B)(6) 2010 (SEE NUMBER ABOVE), IN WHICH ZIMMER STRONGLY RECOMMENDS THE USE OF A DROP DOWN STEM EXTENSION IN CONJUNCTION WITH THE BASEPLATE. THE DEVICE IN QUESTION WAS IMPLANTED PRIOR TO THIS FIELD ACTION.
IT IS REPORTED THAT THE PT WAS REVISED DUE TO LOOSENING OF THE TIBIAL BASEPLATE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NEXGEN MIS STEMMED TIBIAL COMPONENT | KNEE PROSTHESIS | JWH | ZIMMER, INC. | 60854135 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR | Required Intervention |