FDA Adverse Event Injury Summary report: N

LOGIC TIBIA PS MOD INSRT SZ 2.5 11MM

MDR report key: 19297748 · Received May 13, 2024

Report

Report Number
1038671-2024-01201
Event Type
Injury
Date Received
May 13, 2024
Date of Event
April 23, 2024
Report Date
March 17, 2025
Manufacturer
EXACTECH, INC.
Product Code
JWH
UDI-DI
10885862001771
PMA / PMN Number
K093360
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL AND/OR CORRECTED INFORMATION. THE FOLLOWING SECTIONS WERE UPDATED/CORRECTED: D6B, H6. MDR SECTION CODES UPDATED/CORRECTED: B, C, D, E, G. THE REVISION REPORTED WAS LIKELY THE RESULT OF PROSTHESIS WEAR. THE WEAR MAY HAVE RESULTED FROM HIGH IN VIVO LOADING ON THE POSTERIOR ASPECT OF THE TIBIAL INSERT AND/OR THIRD BODY WEAR, BUT THE POTENTIAL ROOT CAUSES COULD NOT BE CONFIRMED. POTENTIAL CONTRIBUTIONS OF USER AND PATIENT-RELATED CONSIDERATIONS TO THE EVENT COULD NOT BE ASSESSED AS THE DEVICES WERE NOT AVAILABLE FOR EVALUATION AND RADIOGRAPHS AND RELEVANT CLINICAL INFORMATION WERE NOT PROVIDED. IF ANY FURTHER INFORMATION IS OBTAINED THAT WOULD CHANGE OR ALTER ANY INFORMATION PROVIDED, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Additional Manufacturer Narrative · 0

PENDING INVESTIGATION. D10: 132652 620-00-02 - PLATELET RICH PLASMA KIT WITH SPRAY TIPS (B)(6) 02-010-01-0225 - LOGIC FEMORAL PS CEM LEFT SZ 2.5 (B)(6) 02-012-45-2525 - LGC TIBIAL FIT TRAY CEM SZ 2.5F / 2.5T (B)(6) 200-02-32 - THREE PEG PATELLA 32MM.

Description of Event or Problem · 0

AS REPORTED, THE PATIENT HAD AN INITIAL LEFT TKA ON (B)(6) 2015. THE PATIENT WAS REVISED ON (B)(6) 2024 DUE TO A RECALLED POLY. THE POLY WAS SWAPPED OUT FOR A NON-RECALLED ONE. THERE WAS NO REPORTED BREAKAGE OF A DEVICE OR SURGICAL DELAY/PROLONGATION. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
846225 LOGIC TIBIA PS MOD INSRT SZ 2.5 11MM PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL JWH EXACTECH, INC. 10885862001771

Patients

Seq Age Sex Outcome Treatment
1 81 YR Female Required Intervention