FDA Adverse Event Injury Summary report: N

LOGIC FEMORAL PS CEM RIGHT SZ 3

MDR report key: 19296367 · Received May 13, 2024

Report

Report Number
1038671-2024-01190
Event Type
Injury
Date Received
May 13, 2024
Date of Event
April 22, 2024
Report Date
December 16, 2024
Manufacturer
EXACTECH, INC.
Product Code
JWH
PMA / PMN Number
K033883
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

SECTION H10: (H3) PENDING EVALUATION (D10) CONCOMITANT DEVICE(S): 2013998, 02-012-35-3015 - LOGIC TIBIA PS MOD INSRT SZ 3 15MM, 2025335, 02-012-45-3020 - LGC TIBIAL FIT TRAY CEM SZ 3F / 2T, 2184803, 200-02-32 - THREE PEG PATELLA 32MM, 2191764, 02-012-45-3020 - LGC TIBIAL FIT TRAY CEM SZ 3F / 2T, 2195603, 02-010-01-0330 - LOGIC FEMORAL PS CEM RIGHT SZ 3 ,2223753, 200-02-32 - THREE PEG PATELLA 32MM, 2238446, 02-010-01-0230 - LOGIC FEMORAL PS CEM LEFT SZ 3.

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL AND/OR CORRECTED INFORMATION. THE REVISION REPORTED WAS LIKELY THE RESULT OF AN INSUFFICIENT BOND BETWEEN THE FEMORAL COMPONENT AND BONE WHICH LED TO ASEPTIC (NON-INFECTED) FEMORAL LOOSENING AND PAIN AFTER 12 YEARS OF IMPLANTATION. HOWEVER, POTENTIAL CONTRIBUTIONS FROM PROSTHESIS WEAR, AND/OR PATIENT-RELATED CONDITIONS TO THE REPORTED EVENT CANNOT BE CONFIRMED AS THE DEVICES WERE NOT AVAILABLE FOR EVALUATION AND LIMITED INFORMATION WAS PROVIDED. IF ANY FURTHER INFORMATION IS OBTAINED THAT WOULD CHANGE OR ALTER ANY INFORMATION PROVIDED, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Description of Event or Problem · 0

IT WAS REPORTED THAT THIS 63 Y/O FEMALE PATIENT'S RIGHT KNEE WAS REVISED APPROXIMATELY 12 YEARS POST OP INITIAL SURGERY. PATIENT COMPLAINED OF PAIN. LOOSE FEMUR AND RECALLED POLY. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. DEVICES ARE NOT RETURNING, DUE TO RECALL HOSPITAL WILL NOT RELEASE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
449296 LOGIC FEMORAL PS CEM RIGHT SZ 3 PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL JWH EXACTECH, INC. UNK

Patients

Seq Age Sex Outcome Treatment
1 63 YR Female Required Intervention SEE H10