FDA Adverse Event Injury Summary report: N

LOGIC TIBIA IMPLANT PSC INSERT, SZ 4, 9MM

MDR report key: 19094661 · Received April 12, 2024

Report

Report Number
1038671-2024-00813
Event Type
Injury
Date Received
April 12, 2024
Date of Event
November 7, 2022
Report Date
October 28, 2024
Manufacturer
EXACTECH, INC.
Product Code
JWH
UDI-DI
10885862173638
PMA / PMN Number
K110547
Removal / Correction Number
Z-0021-2022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

CONCOMITANTS:(B)(6) 02-010-01-0340 - LOGIC FEMORAL PS CEM RIGHT SZ 4 (B)(6) 02-012-45-4040 - LGC TIBIAL FIT TRAY CEM SZ 4F / 4T (B)(6) 200-02-35 - THREE PEG PATELLA 35MM (B)(6) 203-96-03 - (11-2216) SAW BLADE NEW STRYKER (.050) (B)(6) 203-96-62 - SAWBLADES (B)(6) 90X22, 1.19MM STKR 5/6/7. THESE DEVICES ARE USED FOR TREATMENTS, NOT DIAGNOSIS. THERE IS NO OTHER INFORMATION AVAILABLE.

Additional Manufacturer Narrative · 0

H6: CORRECTED THE FOLLOWING: HEALTH EFFECT - CLINICAL CODE, TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. MANUFACTURER NARRATIVE UPDATED: THE REASON FOR THE REVISION REPORTED IN INCIDENT CANNOT BE CONFIRMED FROM THE INFORMATION PROVIDED BUT MAY BE THE RESULT OF PROSTHESIS WEAR OR DUE TO INCLUSION OF THE POLYETHYLENE IN THE PACKAGING RECALL. ADDITIONALLY, A CONTRIBUTING FACTOR TO THE REPORTED WEAR MAY HAVE BEEN THE RESULT OF BEING PACKAGED IN A NON-CONFORMING VACUUM BAG FOR MORE THAN FIVE YEARS. POTENTIAL CONTRIBUTIONS OF USER AND PATIENT-RELATED CONSIDERATIONS TO THE EVENT COULD NOT BE ASSESSED AS THE DEVICES WERE NOT AVAILABLE FOR EVALUATION AND IMAGES, RADIOGRAPHS, AND RELEVANT CLINICAL INFORMATION WERE NOT PROVIDED.

Description of Event or Problem · 0

IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT A PATIENT HAD A RIGHT TOTAL KNEE ARTHROPLASTY ON (B)(6) 2020, AND THEN EXPERIENCED REVISION SURGICAL PROCEDURE ON (B)(6) 2022 APPROXIMATELY 2 YEARS AND 4 MONTHS AFTER INITIAL IMPLANT. NO IMAGES WERE PROVIDED. THERE IS NO OTHER INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1401532 LOGIC TIBIA IMPLANT PSC INSERT, SZ 4, 9MM PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL JWH EXACTECH, INC. 10885862173638

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention SEE H10.