FDA Adverse Event Injury Summary report: N

DARWIN CR/SB PS TB IN 10M/SZ 4

MDR report key: 1903331 · Received November 16, 2010

Report

Report Number
1818910-2010-08551
Event Type
Injury
Date Received
November 16, 2010
Date of Event
October 20, 2010
Report Date
October 20, 2010
Manufacturer
DEPUY-RAYNHAM, A DIVISION OF DEPUY ORTHOPAEDICS, INC.
Product Code
JWH
PMA / PMN Number
K950010
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT ASSOCIATED WITH THIS REPORTED EVENT WAS NOT RETURNED FOR EXAMINATION. A SEARCH OF THE COMPLAINT DATABASE DID NOT SHOW ANY OTHER REPORTS AGAINST THE PRODUCT LOT CODE. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS ABOUT THE REPORTED EVENT WITHOUT THE PRODUCT TO EXAMINE. PROVIDED INFO STATED THE PRODUCT WAS NOT SUSPECTED OF FAILING TO MEET SPECIFICATIONS OR BEING A CONTRIBUTING FACTOR TO THE EVENT. NO EVIDENCE WAS FOUND SUGGESTING PRODUCT ERROR WAS A CONTRIBUTING FACTOR AND THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFO BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.

Description of Event or Problem · 1

PT WAS REVISED TO ADDRESS PAIN DUE TO FEMORAL LOOSENING AT THE CEMENT/IMPLANT INTERFACE (COMPETITOR'S CEMENT WAS USED IN THE PRIMARY SURGERY). POLY WEAR OF THE INSERT AND OSTEOLYSIS WERE ALSO REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DARWIN CR/SB PS TB IN 10M/SZ 4 87JWH JWH DEPUY-RAYNHAM, A DIVISION OF DEPUY ORTHOPAEDICS, INC. NA 631DE

Patients

Seq Age Sex Outcome Treatment
1 56 YR Required Intervention