ARISTOTLE 14
Report
- Report Number
- 3009546466-2023-00006
- Event Type
- Injury
- Date Received
- November 15, 2023
- Date of Event
- October 10, 2023
- Report Date
- November 8, 2023
- Manufacturer
- SCIENTIA VASCULAR
- Product Code
- MOF
- UDI-DI
- 00818075010011
- PMA / PMN Number
- K173235
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
FROM THE REVIEW OF THE MANUFACTURING LOT HISTORY RECORD, THERE IS NO INDICATION THAT THE BREAK OCCURRED DUE TO A MANUFACTURING DEFECT, AND IT DETERMINED THE LOT MET THE RELEASING CRITERIA. THERE IS NO WAY TO CONFIRM EVERY STEP TAKEN THAT COULD HAVE LED THE GUIDEWIRE TO BREAK, BUT FROM THE INFORMATION COLLECTED IN THIS INVESTIGATION, THE GUIDEWIRE APPEARS TO HAVE BENT DURING THE PROCEDURE AND THE BEND MAY HAVE BEEN FATIGUED DURING USE. THE BEND MAY HAVE BEEN FURTHER STRESSED WHEN THE PHYSICIAN TORQUED THE DEVICE AFTER IT WAS STUCK UNTIL IT BROKE. AN ATTEMPT TO SUBMIT THE EMDR WAS PERFORMED ON 08-NOV-2023, AND ON 09-NOV-2023. HOWEVER, THE ACCOUNT CERTIFICATE NEEDED TO BE PROCESSED BY THE ESG HELP DESK AND BE READY FOR USE. THIS ACTIVITY LED TO A SUBMISSION DELAY THAT WENT OVER THE 30 DAYS FDA REQUIREMENT FOR THIS TYPE OF EMDR. THE ESG HELP DESK SUGGESTED THAT THIS INFORMATION BE ENTER IN THIS TAB.
ON THE 10TH OF OCTOBER 2023, SCIENTIA VASCULAR WAS NOTIFIED THAT AN ARISTOTLE 14 GUIDEWIRE (CATALOG # A14-200-002 LN:022981) BROKE DURING A PROCEDURE. THE EVENT WAS DESCRIBED AS FOLLOWS: "THE GUIDEWIRE GOT STUCK. PHYSICIAN CONTINUED TO TORQUE THE DEVICE, COULD NOT GET THE GUIDEWIRE OUT, GUIDEWIRE ENDED UP BREAKING INSIDE THE PATIENT, HOWEVER THE DOCTOR WAS ABLE TO USE A SNARE TO REMOVE THE GUIDEWIRE FROM THE PATIENT. THERE WAS NO PATIENT INJURY REPORTED FOR THIS EVENT."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 316066 | ARISTOTLE 14 | GUIDEWIRE | MOF | SCIENTIA VASCULAR | ARISTOTLE 14 | 022981 | 00818075010011 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Required Intervention | SL-10 MICROCATHETER |