FDA Adverse Event Injury Summary report: N

LOGIC CR TIB INSERT SLOPE+, SZ 2, 11MM

MDR report key: 18119248 · Received November 13, 2023

Report

Report Number
1038671-2023-02753
Event Type
Injury
Date Received
November 13, 2023
Date of Event
October 31, 2017
Report Date
October 21, 2024
Manufacturer
EXACTECH, INC.
Product Code
JWH
UDI-DI
10885862174291
PMA / PMN Number
K111400
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

D10: CONCOMITANTS: 02-012-45-2020 - LGC TIBIAL FIT TRAY CEM SZ 2F / 2T, 5099125. 02-012-45-2515 - LGC TIBIAL FIT TRAY CEM SZ 2.5F / 1.5T, 4213359. 02-020-13-0225 - TRULIANT CR CEM FEM CR CEM LEFT SZ 2.5, 4909303. 02-020-13-0320 - TRULIANT CR CEM FEM CR CEM RIGHT SZ 2, 5045053. 200-07-29 - ADVANCED PATELLA 29M 3 PEG IMPLANT, 5015351. 200-07-29 - ADVANCED PATELLA 29M 3 PEG IMPLANT, 5015359. PENDING INVESTIGATION.

Additional Manufacturer Narrative · 0

THE REASON FOR THE REVISION REPORTED CANNOT BE CONFIRMED FROM THE INFORMATION PROVIDED BUT MAY BE THE RESULT OF PROSTHESIS WEAR AND INSTABILITY OR DUE TO INCLUSION OF THE POLYETHYLENE IN THE PACKAGING RECALL. ADDITIONALLY, A CONTRIBUTING FACTOR TO THE REPORTED WEAR MAY HAVE BEEN THE RESULT OF BEING PACKAGED IN A NON-CONFORMING VACUUM BAG FOR MORE THAN FIVE YEARS. POTENTIAL CONTRIBUTIONS OF USER AND PATIENT-RELATED CONSIDERATIONS TO THE EVENT COULD NOT BE ASSESSED AS THE DEVICES WERE NOT AVAILABLE FOR EVALUATION AND IMAGES AND RADIOGRAPHS WERE NOT PROVIDED. H6: CORRECTED MEDICAL DEVICE, COMPONENT, AND INVESTIGATION CLINICAL CODES.

Description of Event or Problem · 0

AS REPORTED VIA LEGAL DOCUMENTATION THE PATIENT HAD A LEFT KNEE REPLACEMENT ON (B)(6) 2017. THIS DEVICE HAS NOT BEEN EXPLANTED. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE. THE PATIENT HAS FILED A SHORT-FORM COMPLAINT IN A MULTIDISTRICT LITIGATION TITLED IN RE: EXACTECH POLYETHYLENE ORTHOPEDIC PRODUCTS LIABILITY LITIGATION, MDL NO. 3044 (NGG) (MMH), AND PENDING IN THE EASTERN DISTRICT OF NEW YORK. PER THE TRANSFER ORDER CREATING THIS MULTIDISTRICT LITIGATION, ¿PLAINTIFFS¿ALLEGE THAT THEIR KNEE OR HIP REPLACEMENT DEVICES¿FAILED PREMATURELY BECAUSE OF DEGRADATION OF THE DEVICE¿S POLYETHYLENE COMPONENT, WHICH RESULTED IN THE PREMATURE REMOVAL (OR PLANNED REMOVAL) OF THE PROSTHESIS AT ISSUE.¿ BECAUSE THE PATIENT HAS FILED A SUIT IN THIS MULTIDISTRICT LITIGATION, THE PATIENT APPEARS TO ALLEGE THAT THE PATIENT WAS INJURED AS A RESULT OF PREMATURE WEAR OF AN EXACTECH POLYETHYLENE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2161989 LOGIC CR TIB INSERT SLOPE+, SZ 2, 11MM PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL JWH EXACTECH, INC. UNK 10885862174291

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention SEE H10.