FDA Adverse Event Injury Summary report: N

OPTETRAK LOGIC

MDR report key: 16999584 · Received May 24, 2023

Report

Report Number
1038671-2023-01129
Event Type
Injury
Date Received
May 24, 2023
Date of Event
December 12, 2022
Report Date
March 25, 2026
Manufacturer
EXACTECH, INC.
Product Code
JWH
UDI-DI
10885862001801
PMA / PMN Number
K033883
Removal / Correction Number
Z-0021-2022
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
501

Narratives

Additional Manufacturer Narrative · 0

H6: THE REVISION REPORTED MAY HAVE BEEN DUE TO THIRD BODY WEAR, PATIENT-RELATED CONDITIONS, INSTABILITY, AND/OR MALALIGNMENT BETWEEN THE FEMORAL AND TIBIAL COMPONENTS, WHICH LED TO PROSTHESIS WEAR. HOWEVER, THIS CANNOT BE CONFIRMED AS THE DEVICES WERE NOT AVAILABLE FOR EVALUATION AND IMAGES OF THE EXPLANTED DEVICE/RADIOGRAPHS WERE NOT PROVIDED.

Additional Manufacturer Narrative · 0

D2B: PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER. D10: CONCOMITANTS: 02-010-01-0230 - FEMUR PS CEM.Nº3 IZQ, 2968030; 02-012-45-3030 - BANDEJA TIBIAL LOGIC FIT 3F/3T, 2966489. ADDITIONAL INFORMATION, INCLUDING THE PRODUCT INVESTIGATION, WILL BE SUBMITTED WITHIN 30 DAYS OF RECEIPT.

Description of Event or Problem · 0

DURING THE WEEK OF APRIL 17-25, REPRESENTATIVES OF EXACTECH, INC. AND EXACTECH SPAIN CONDUCTED AN INPERSON SITE VISIT WITH (B)(6) HOSPITAL. AS A RESULT OF THAT SITE VISIT AND AT EXACTECH'S REQUEST, THE HOSPITAL PROVIDED A RETROSPECTIVE EXCEL LISTING OF REVISION RELATED TO POLYETHYLENE WEAR THAT HAVE BEEN PERFORMED SINCE 2010. THESE CASES HAVE NOT PREVIOUSLY BEEN REPORTED TO EXACTECH AS COMPLAINTS. THIS PATIENT (B) (6) WAS IMPLANTED WITH A 02-012-35-3009-INSERTO TIBIAL LOGIC PS 3, 9 MM, SERIAL NUMBER (B)(6) ON (B)(6) 2014. THE INSERT WAS MANUFACTURED ON 2/5/2014 AND WAS PACKAGED IN A VACUUM BAG WITH NO EVOH. THE INSERT WAS MANUFACTURED ON 2/5/2014 AND WAS 0.3 YEARS SHELF AGE AT THE TIME OF IMPLANT. ON (B)(6) 2022, APPROXIMATELY 8.0 YEARS POST IMPLANTATION, THE PATIENT UNDERWENT REVISION FOR WEAR OF THE POLYETHYLENE. NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
809102 OPTETRAK LOGIC SEE H10 JWH EXACTECH, INC. LOGIC TIBIA PS MOD INSRT SZ 3 9MM UNK 10885862001801

Patients

Seq Age Sex Outcome Treatment
1 Unknown Required Intervention