CERELINK ICP PROBE BASIC KIT
Report
- Report Number
- 3013886523-2023-00122
- Event Type
- Injury
- Date Received
- April 25, 2023
- Date of Event
- April 5, 2023
- Report Date
- July 7, 2023
- Manufacturer
- INTEGRA LIFESCIENCES SWITZERLAND SAR
- Product Code
- GWM
- PMA / PMN Number
- K173192
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THE MICROSENSOR (ID 826851) WAS RETURNED FOR EVALUATION. DEVICE HISTORY RECORD (DHR) - LOT 6202214, SN (B)(6) MET SPECIFICATIONS WHEN RELEASED. FAILURE ANALYSIS - THE ISSUE OF THE COMPLAINT HAS BEEN CONFIRMED: - CATHETER NICKED 14CM FROM SENSOR TIP EXPOSING INTERNAL WIRES; AND TAPE BUNCHED UP AROUND CATHETER - CATHETER COVERED IN GAUZE IN TAPE - CATHETER TWISTED INTO A LOOP IN THE MIDDLE OF CATHETER - CERELINK MONITOR READ ¿ACCEPTABLE¿; ICP EXPRESS = 520 - NO TESTING WAS POSSIBLE, DUE TO THE INTERNAL WIRES EXPOSED. THE POSSIBLE ROOT CAUSE FOR THIS ISSUE REPORTED BY THE CUSTOMER COULD BE DUE TO A MISHANDLING OF THE CATHETER.
AN INVESTIGATION HAS BEEN INITIATED BASED ON THE REPORTED INFORMATION. UPON COMPLETION OF THE INVESTIGATION, A FOLLOW-UP REPORT WILL BE SUBMITTED.
THIS IS 1 OF 2 REPORTS LINKED TO MFG REPORT NUMBERS: 3013886523-2023-00123. A FACILITY REPORTED A MICROSENSOR (ID 826851-SERIAL (B)(6) WAS SHOWING NEGATIVE NUMBERS ON THE ICP EXPRESS MONITOR. THE FIRST MICROSENSOR WAS IMPLANTED ON (B)(6) 2023 AND SUDDENLY STARTED TO READ NUMBERS IN THE RANGE OF -8 TO -5. THEREFORE, THE MICROSENSOR WAS REMOVED ON (B)(6) 2023 AND WAS REPLACED WITH ANOTHER MICROSENSOR (ID 826851-SERIAL (B)(6). THE REPLACED MICROSENSOR DROPPED TO -8 AND WAS ALSO REMOVED. THE TIMEFRAME BETWEEN IMPLANTATION AND EXPLANTATION DATE WAS 24 HOURS EACH FOR BOTH SENSORS. THE PATIENT RECOVERED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1188314 | CERELINK ICP PROBE BASIC KIT | ICP MICRSOSENSORS - CERELINK | GWM | INTEGRA LIFESCIENCES SWITZERLAND SAR |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | ICP EXPRESS |