FDA Adverse Event Malfunction Summary report: N

GMK-SPHERE TIBIAL INSERT FIXED SPHERE FLEX SIZE 4/10 MM R

MDR report key: 15931582 · Received December 7, 2022

Report

Report Number
3005180920-2022-00920
Event Type
Malfunction
Date Received
December 7, 2022
Date of Event
November 11, 2022
Report Date
December 7, 2022
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JWH
UDI-DI
07630030826542
PMA / PMN Number
K121416
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 18 NOVEMBER 2022 LOT 143392: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 03-JUL-2014. EXPIRATION DATE: 2019-MAY-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. SINCE RELEASE TO DATE, ALL ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT. CLINICAL EVALUATION PERFORMED BY MEDACTA MEDICAL AFFAIRS DIRECTOR: PROGRESSIVE UNSCREWING OF THE INSERT FIXATION SCREW IN A TKA IMPLANTED CORRECTLY 8 YEARS BEFORE. THE TORQUE-LIMITING DEVICE HAD NOT BEEN USED AT THE TIME. THE SCREW IS PRESENTLY IN A POSITION WHERE NO DAMAGE IS TO BE EXPECTED, UNLESS THE PATIENT FEELS IT - BUT THIS SEEMS NOT TO BE THE CASE. OF COURSE THE SCREW COULD MIGRATE FURTHER AND BE LOCATED IN A POSITION WHERE REMOVAL BECOMES NECESSARY. THE REASONS FOR SELF-UNSCREWING OF THE FIXATION SCREW CANNOT BE DETERMINED WITH CERTAINTY: ONE POSSIBILITY, THAT COULD BE REPRODUCED IN THE LABORATORY, IS INSUFFICIENT TIGHTENING FORCE, BUT OTHER UNKNOWN FACTORS MAY ALSO HAVE PLAYED A ROLE.

Description of Event or Problem · 0

ABOUT 8 YEARS AFTER THE PRIMARY SURGERY, THE INSERT SCREW UNSCREWED. AT THIS TIME, AN ADDITIONAL/REVISION SURGERY IS NOT PLANNED, THE PATIENT DOES NOT FEEL ANY PAIN OR DISCOMFORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1470263 GMK-SPHERE TIBIAL INSERT FIXED SPHERE FLEX SIZE 4/10 MM R KNEE TIBIAL INSERT (FIXATION SCREW) JWH MEDACTA INTERNATIONAL SA 02.12.0410FR 143392 07630030826542

Patients

Seq Age Sex Outcome Treatment
1 83 YR Male Other