FDA Adverse Event Injury Summary report: N

GMK-SPHERE TIBIAL TRAY FIXED CEMENTED SIZE 5 L

MDR report key: 15822855 · Received November 18, 2022

Report

Report Number
3005180920-2022-00869
Event Type
Injury
Date Received
November 18, 2022
Date of Event
October 28, 2022
Report Date
November 11, 2022
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JWH
UDI-DI
07630030819926
PMA / PMN Number
K090988
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON (B)(6) 2022 LOT 2113305: 54 ITEMS MANUFACTURED AND RELEASED ON (B)(6) 2021. EXPIRATION DATE: 2026-12-08. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 48 ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. ADDITIONAL IMPLANTS INVOLVED: BATCH REVIEW PERFORMED ON (B)(6) 2022 ON GMK-SPHERE 02.12.0510FL TIBIAL INSERT FIXED SPHERE FLEX SIZE 5/10 MM L (K121416) LOT. 2112472 LOT 2112472: 125 ITEMS MANUFACTURED AND RELEASED ON (B)(6) 2021. EXPIRATION DATE: 2026-10-11. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 119 ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. BATCH REVIEW PERFORMED ON (B)(6) 2022 ON GMK-SPHERE 02.12.0006L FEMORAL COMPONENT SPHERE CEMENTED SIZE 6 L (K121416) LOT. 2109966 LOT 2109966: 20 ITEMS MANUFACTURED AND RELEASED ON (B)(6) 2021. EXPIRATION DATE: 2026-11-02. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 19 ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW.

Description of Event or Problem · 0

AT ABOUT 6 MONTHS AFTER THE PRIMARY, THE PATIENT CAME IN REPORTING PATELLA PAIN AND INSTABILITY. THE SURGEON RESURFACED PATELLA AND SAW THAT THE PATIENT WAS VERY UNSTABLE IN FLEXION. THE SURGEON CONVERTED THE PATIENT TO A HINGE TO STABILIZE THE KNEE. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
228260 GMK-SPHERE TIBIAL TRAY FIXED CEMENTED SIZE 5 L KNEE TIBIAL TRAY CEMENTED JWH MEDACTA INTERNATIONAL SA 02.07.1205L 2113305 07630030819926

Patients

Seq Age Sex Outcome Treatment
1 Male Required Intervention