TRULIANT PS CEM FEM PS CEM LEFT SZ 4
Report
- Report Number
- 1038671-2022-01174
- Event Type
- Injury
- Date Received
- September 27, 2022
- Date of Event
- September 1, 2022
- Report Date
- March 2, 2026
- Manufacturer
- EXACTECH, INC.
- Product Code
- JWH
- UDI-DI
- 10885862314451
- PMA / PMN Number
- K170240
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
PENDING EVALUATION. CONCOMITANT DEVICE(S): S035099, 02-022-35-4009 - TRULIANT TIB IMP PS INSERT SZ 4 9MM; 6222387, 02-022-45-4040 - TRULIANT TIB FIT TRAY CEM SZ 4F / 4T; 6190986, 200-07-35 - ADVANCED PATELLA 35MM 3 PEG IMPLANT.
SECTION H10: (H3) THE REVISION REPORTED WAS LIKELY THE RESULT OF AN INSUFFICIENT BOND BETWEEN THE FEMORAL COMPONENT AND THE BONE, WHICH LED TO ASEPTIC (NON-INFECTED) FEMORAL LOOSENING AND PAIN. HOWEVER, THIS CANNOT BE CONFIRMED AS THE DEVICE WAS NOT RETURNED FOR EVALUATION.
AS REPORTED, APPROXIMATELY 3 YEARS POST OP THE INITIAL LEFT TKA, THIS 75 Y/O MALE PATIENT WAS REVISED DUE TO A LOOSE FEMUR. PATIENT COMPLAINED OF PAIN. DUE TO RECALL SURGEON USED A COMPETITOR FOR REVISION. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. PRODUCT NOT RETURNING, DUE TO RECALL HOSPITAL WILL NOT RELEASE THE IMPLANTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2304271 | TRULIANT PS CEM FEM PS CEM LEFT SZ 4 | PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL | JWH | EXACTECH, INC. | 02-020-11-0240 | UNK | 10885862314451 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | Male | Required Intervention |