FDA Adverse Event Malfunction Summary report: N

GII PATELLAR REAM GUIDE

MDR report key: 13486217 · Received February 8, 2022

Report

Report Number
1020279-2022-00585
Event Type
Malfunction
Date Received
February 8, 2022
Date of Event
January 14, 2022
Report Date
April 1, 2022
Manufacturer
SMITH & NEPHEW, INC.
Product Code
JWH
UDI-DI
03596010214911
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

INTERNAL COMPLAINT REFERENCE CASE-(B)(4).

Additional Manufacturer Narrative · 0

H10: THE DEVICE WAS NOT RETURNED FOR EVALUATION AND THE REPORTED EVENT COULD NOT BE CONFIRMED. A REVIEW OF COMPLAINT HISTORY OF THE PREVIOUS 12 MONTHS REVEALED SIMILAR EVENTS FOR THE LISTED DEVICE, THIS FAILURE MODE WILL BE MONITORED FOR FUTURE COMPLAINTS FOR ANY NECESSARY CORRECTIVE ACTIONS. A REVIEW OF THE RISK MANAGEMENT FILE REVEALED THIS FAILURE MODE WAS PREVIOUSLY IDENTIFIED. THE ANTICIPATED RISK LEVEL IS STILL ADEQUATE. A HISTORICAL REVIEW CONCLUDED THAT THERE ARE NO PRIOR ACTIONS RELATED TO THIS PRODUCT AND EVENT. AT THIS TIME, WE HAVE NO REASON TO SUSPECT THAT THE PRODUCT FAILED TO MEET ANY PRODUCT SPECIFICATIONS AT THE TIME OF MANUFACTURE. THIS DEVICE IS A REUSABLE INSTRUMENT THAT CAN BE EXPOSED TO NUMEROUS SURGERIES. DAMAGE FROM PROLONGED USE, MISUSE OR ROUGH HANDLING ARE LIKELY PROBABLE CAUSES OF THE REPORTED EVENT. WE RECOMMEND THAT ALL REUSABLE INSTRUMENTS BE ROUTINELY INSPECTED FOR WEAR AND DAMAGE AND REPLACED AS NECESSARY. THE CONTRIBUTION OF THE DEVICE TO THE REPORTED EVENT COULD NOT BE CORROBORATED. BASED ON THIS INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. WITHOUT THE RETURN OF THE ACTUAL PRODUCT INVOLVED, OUR INVESTIGATION COULD NOT PROCEED. SHOULD THE DEVICE OR ADDITIONAL INFORMATION BE RECEIVED, THE COMPLAINT WILL BE REOPENED. NO FURTHER INVESTIGATION IS WARRANTED FOR THIS COMPLAINT; HOWEVER, WE WILL CONTINUE TO MONITOR FOR FUTURE COMPLAINTS AND INVESTIGATE AS NECESSARY. WE CONSIDER THIS INVESTIGATION CLOSED. INTERNAL COMPLAINT REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT, DURING A TKA SURGERY, THE GOLD KOB OF A GII PATELLAR REAM GUIDE WAS NOT LOCKING CLAMP INTO PLACE, IT COMES LOOSE. SURGERY WAS PERFORMED, WITHOUT ANY DELAY, USING A SN BACK-UP DEVICE. PATIENT WAS NOT HARMED AS A CONSEQUENCE OF THIS ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1850354 GII PATELLAR REAM GUIDE PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL JWH SMITH & NEPHEW, INC. 5575793 03596010214911

Patients

Seq Age Sex Outcome Treatment
1 Unknown