FDA Adverse Event Injury Summary report: N

GMK-PRIMARY TIBIAL TRAY FIXED CEMENTED SIZE 3 L

MDR report key: 13060403 · Received December 22, 2021

Report

Report Number
3005180920-2021-01002
Event Type
Injury
Date Received
December 22, 2021
Date of Event
November 29, 2021
Report Date
January 21, 2022
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JWH
UDI-DI
07630030819889
PMA / PMN Number
K090988
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS DIRECTOR AT 9 YEARS AFTER PRIMARY TKA BOTH COMPONENTS MOBILIZE QUITE SIGNIFICANTLY, WHILE THE PATELLA REMAINS WELL FIXED. NO TRAUMATIC EVENT IS REPORTED, AND NO INFECTION. THE REASONS FOR SUCH A SUDDEN AND MACROSCOPIC LOOSENING ARE NOT KNOWN AND CANNOT BE DETERMINED WITH THE INFORMATION AT HAND. IDIOPATHIC ASEPTIC LOOSENING IS A POSSIBLE ADVERSE EVENT FOLLOWING TKA, REPORTED IN LITERATURE. THERE IS NO REASON TO SUSPECT FAULTY DEVICES, WITH THE INFORMATION AT HAND.

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 09 DECEMBER 2021: LOT 122122: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 09-AUG-2012. EXPIRATION DATE: 2017-JUN-30. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT SINCE 1-1-2017. ADDITIONAL COMPONENT INVOLVED: GMK-PRIMARY 02.07.2204L FEMUR PS CEMENTED SIZE 4 L (K090988) LOT. 112990. LOT 112990: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 06-OCT-2011. EXPIRATION DATE: 2016-AUG-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT SINCE 1-1-2017.

Description of Event or Problem · 0

REVISION SURGERY AT 9 YEARS AND 1 MONTH AFTER PRIMARY DUE TO FEMUR AND TIBIAL IMPLANT MOBILIZATION. ALL IMPLANTS REVISED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1964145 GMK-PRIMARY TIBIAL TRAY FIXED CEMENTED SIZE 3 L KNEE TIBIAL BASEPLATE CEMENTED JWH MEDACTA INTERNATIONAL SA 02.07.1203L 122122 07630030819889

Patients

Seq Age Sex Outcome Treatment
1 76 YR Female Required Intervention