FDA Adverse Event Malfunction Summary report: N

NEXGEN CR-FLEX PCT FEM G-R MINUS

MDR report key: 12960174 · Received December 8, 2021

Report

Report Number
3007963827-2021-00313
Event Type
Malfunction
Date Received
December 8, 2021
Date of Event
November 11, 2021
Report Date
January 7, 2022
Manufacturer
ZIMMER ORTHOPAEDIC MFG. LTD.
Product Code
JWH
UDI-DI
00889024207080
PMA / PMN Number
K173057
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(4). PRODUCT HAS BEEN RETURNED TO ZIMMER BIOMET, AND THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION IS COMPLETED, A SUPPLEMENTAL MEDWATCH 3500A WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. THE RESULTS OF THE INVESTIGATION ARE AS FOLLOWS: VISUAL INSPECTION OF THE RETURNED PRODUCT FOUND THE CARTON EXHIBITS SOME WEAR BUT NO CREASES OR DENTS. BOTH INNER CAVITIES ARE CONFIRMED CRACKED/DAMAGED. STERILITY HAS BEEN COMPROMISED. REVIEW OF THE DEVICE HISTORY RECORD(S) IDENTIFIED NO DEVIATIONS OR ANOMALIES DURING MANUFACTURING. THE CONDITION OF THE DEVICE WHEN IT LEFT ZIMMER BIOMET IS CONFORMING TO SPECIFICATION. THE ROOT CAUSE OF THE REPORTED EVENT CAN BE ATTRIBUTED TO TRANSIT DAMAGE. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE INNER AND OUTER PLASTIC PACKAGING WAS CRACKED. IT IS THOUGHT TO BE MOST LIKELY DUE TO THE IMPLANTS BEING WRAPPED TOO TIGHTLY IN THE PACKAGING BUNDLE. NO CONSEQUENCES OR IMPACT TO THE PATIENT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1859517 NEXGEN CR-FLEX PCT FEM G-R MINUS PROSTHESIS, KNEE JWH ZIMMER ORTHOPAEDIC MFG. LTD. N/A 63954628 00889024207080

Patients

Seq Age Sex Outcome Treatment
1 43 YR Male