FDA Adverse Event Malfunction Summary report: N

2 UNIT RED CELL SET W/RC2H FIL

MDR report key: 12711245 · Received October 28, 2021

Report

Report Number
1219343-2021-00136
Event Type
Malfunction
Date Received
October 28, 2021
Date of Event
September 28, 2021
Report Date
October 27, 2021
Manufacturer
HAEMONETICS CORPORATION
Product Code
GKT
UDI-DI
10812747013512
PMA / PMN Number
BK060023
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

ON (B)(6) 2021, HAEMONETICS MANUFACTURING PERFORMED A VISUAL INSPECTION OF THE RETURNED SAMPLE AND IT WAS CONFIRMED THAT THE CONFIGURATION OF THE UNIT IS CORRECT (ALL COMPONENTS ARE CORRECTLY ASSEMBLED). KINKS WERE FOUND THROUGHOUT THE ALL TUBING HOWEVER, SINCE THE RETURNED SAMPLE WAS RECEIVED WITH THE TUBING'S COILED, IT IS NOT POSSIBLE TO CONFIRM IF THESE KINKS WERE GENERATED DUE TO THE WAY THE SAMPLE WAS RETURNED OR FOR OTHER REASON. A VISUAL EVALUATION WAS CONDUCTED AND IT WAS FOUND THAT THE 6 FILTER HOLES ARE NOT BLOCKED. BOTH SIDES OF THE FILTER WERE CHECKED AND THERE ARE NO TRACES OF SOLVENT ON EITHER SIDE. THE MEDIA IS DAMAGED WITH ONE HOLE CLOSE TO THE OUTLET.

Description of Event or Problem · 1

ON (B)(6) 2021, HAEMONETICS WAS NOTIFIED OF WEIGHT CHANGING TOO SLOWLY IN RBC TRANSFER, WEIGHT CHANGING TO SLOWLY IN PLASMA RETURN, AIR DETECTED AT BLAD ALARMS DURING RETURN AND POSSIBLE RBC SPILL DURING DRAW ALARMS UTILIZING A 2 UNIT RED CELL SET.