2 UNIT RED CELL SET W/RC2H FIL
Report
- Report Number
- 1219343-2021-00136
- Event Type
- Malfunction
- Date Received
- October 28, 2021
- Date of Event
- September 28, 2021
- Report Date
- October 27, 2021
- Manufacturer
- HAEMONETICS CORPORATION
- Product Code
- GKT
- UDI-DI
- 10812747013512
- PMA / PMN Number
- BK060023
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- 003
Narratives
ON (B)(6) 2021, HAEMONETICS MANUFACTURING PERFORMED A VISUAL INSPECTION OF THE RETURNED SAMPLE AND IT WAS CONFIRMED THAT THE CONFIGURATION OF THE UNIT IS CORRECT (ALL COMPONENTS ARE CORRECTLY ASSEMBLED). KINKS WERE FOUND THROUGHOUT THE ALL TUBING HOWEVER, SINCE THE RETURNED SAMPLE WAS RECEIVED WITH THE TUBING'S COILED, IT IS NOT POSSIBLE TO CONFIRM IF THESE KINKS WERE GENERATED DUE TO THE WAY THE SAMPLE WAS RETURNED OR FOR OTHER REASON. A VISUAL EVALUATION WAS CONDUCTED AND IT WAS FOUND THAT THE 6 FILTER HOLES ARE NOT BLOCKED. BOTH SIDES OF THE FILTER WERE CHECKED AND THERE ARE NO TRACES OF SOLVENT ON EITHER SIDE. THE MEDIA IS DAMAGED WITH ONE HOLE CLOSE TO THE OUTLET.
ON (B)(6) 2021, HAEMONETICS WAS NOTIFIED OF WEIGHT CHANGING TOO SLOWLY IN RBC TRANSFER, WEIGHT CHANGING TO SLOWLY IN PLASMA RETURN, AIR DETECTED AT BLAD ALARMS DURING RETURN AND POSSIBLE RBC SPILL DURING DRAW ALARMS UTILIZING A 2 UNIT RED CELL SET.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1611711 | 2 UNIT RED CELL SET W/RC2H FIL | 2 UNIT RED CELL SET W/RC2H FIL | GKT | HAEMONETICS CORPORATION | 0832F-00 | 0321042 | 10812747013512 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |