FDA Adverse Event Malfunction Summary report: N

JRNY II CR LKG FEM IMP BUMPER RT

MDR report key: 12468267 · Received September 14, 2021

Report

Report Number
1020279-2021-06825
Event Type
Malfunction
Date Received
September 14, 2021
Date of Event
August 24, 2021
Report Date
September 14, 2021
Manufacturer
SMITH & NEPHEW, INC.
Product Code
JWH
UDI-DI
00885556085233
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE ASSOCIATED DEVICE, INTENDED FOR USE IN TREATMENT, WAS RETURNED AND EVALUATED. A VISUAL INSPECTION OF THE RETURNED JRNY II CR LKG FEM IMP BUMPER RT CONFIRMS THE DEVICE IS BROKE IN HALF AND ONLY ONE PIECE WAS RETURNED. THIS DEVICE ALSO HAS EXCESSIVE WEAR USAGE. A COMPLAINT HISTORY REVIEW FOUND RELATED FAILURES; THIS FAILURE MODE WILL BE MONITORED FOR FUTURE COMPLAINTS AND ASSESSED FOR ANY NECESSARY CORRECTIVE ACTIONS. AT THIS TIME, WE DO NOT HAVE REASON TO SUSPECT THAT THE PRODUCT FAILED TO MEET ANY PRODUCT SPECIFICATIONS AT THE TIME OF MANUFACTURE. THIS DEVICE IS A REUSABLE INSTRUMENT THAT CAN BE EXPOSED TO NUMEROUS SURGERIES. DAMAGE FROM PROLONGED USE, MISUSE OR ROUGH HANDLING ARE LIKELY PROBABLE CAUSES OF THE REPORTED EVENT. WE RECOMMEND THAT ALL REUSABLE INSTRUMENTS BE ROUTINELY INSPECTED FOR WEAR AND DAMAGE AND REPLACED AS NECESSARY. BASED ON THIS INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE COMPLAINT WILL BE REOPENED. NO FURTHER INVESTIGATION IS WARRANTED FOR THIS COMPLAINT; HOWEVER, WE WILL CONTINUE TO MONITOR FOR FUTURE COMPLAINTS AND INVESTIGATE AS NECESSARY.

Description of Event or Problem · 1

IT WAS REPORTED THAT A JRNY II CR LKG FEM IMP BUMPER RT WAS DISCOVERED BROKEN IN HALF. AS THIS WAS NOTICED UPON CLEANING AND STERILIZATION ACTIVITIES, THERE WAS NOT PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1364488 JRNY II CR LKG FEM IMP BUMPER RT PROSTHSISKNEPATLLOFMOROTIBIALSMICONSTRAINDCMNTDPOLYMRMTALPOLYMR JWH SMITH & NEPHEW, INC. 74011857 00885556085233

Patients

Seq Age Sex Outcome Treatment
1