FDA Adverse Event Malfunction Summary report: N

JRNY II CR LKG FEM IMP BUMPER RT

MDR report key: 12211507 · Received July 22, 2021

Report

Report Number
1020279-2021-06022
Event Type
Malfunction
Date Received
July 22, 2021
Date of Event
July 5, 2021
Report Date
August 17, 2021
Manufacturer
SMITH & NEPHEW, INC.
Product Code
JWH
UDI-DI
00885556085233
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

THE ASSOCIATED DEVICE, USED IN TREATMENT, WAS RETURNED AND EVALUATED. A VISUAL INSPECTION OF THE RETURNED JRNY II CR LKG FEM IMP BUMPER RT CONFIRMED THE DEVICE IS BROKEN IN HALF. THE BROKEN PIECE WAS RETURNED WITH THE DEVICE. A COMPLAINT HISTORY REVIEW FOUND RELATED FAILURES; THIS FAILURE MODE WILL BE MONITORED FOR FUTURE COMPLAINTS AND ASSESSED FOR ANY NECESSARY CORRECTIVE ACTIONS. A REVIEW OF THE MANUFACTURING RECORDS DID NOT REVEAL A MANUFACTURING ABNORMALITY THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED INCIDENT. AT THIS TIME, WE DO NOT HAVE REASON TO SUSPECT THAT THE PRODUCT FAILED TO MEET ANY PRODUCT SPECIFICATIONS AT THE TIME OF MANUFACTURE. THIS DEVICE IS A REUSABLE INSTRUMENT THAT CAN BE EXPOSED TO NUMEROUS SURGERIES. DAMAGE FROM PROLONGED USE, MISUSE OR ROUGH HANDLING ARE LIKELY PROBABLE CAUSES OF THE REPORTED EVENT. WE RECOMMEND THAT ALL REUSABLE INSTRUMENTS BE ROUTINELY INSPECTED FOR WEAR AND DAMAGE AND REPLACED AS NECESSARY. BASED ON THIS INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE COMPLAINT WILL BE REOPENED. NO FURTHER INVESTIGATION IS WARRANTED FOR THIS COMPLAINT; HOWEVER, WE WILL CONTINUE TO MONITOR FOR FUTURE COMPLAINTS AND INVESTIGATE AS NECESSARY. WE CONSIDER THIS INVESTIGATION CLOSED.

Description of Event or Problem · 1

IT WAS REPORTED THAT, DURING TKA SURGERY, A JRNY II CR LKG FEM IMP BUMPER RT BROKE WHILE INSTRUMENTS WERE OUTSIDE THE PATIENT. SURGERY WAS RESUMED, WITH THE SAME DEVICE, WITHOUT ANY DELAY. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1107269 JRNY II CR LKG FEM IMP BUMPER RT PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL JWH SMITH & NEPHEW, INC. 74011857 13FBG0026 00885556085233

Patients

Seq Age Sex Outcome Treatment
1