ARCOM SERIES A PAT W/WR STD 31 1 PEG
Report
- Report Number
- 0001825034-2021-01747
- Event Type
- Injury
- Date Received
- June 9, 2021
- Date of Event
- May 11, 2021
- Report Date
- June 8, 2021
- Manufacturer
- ZIMMER BIOMET, INC.
- Product Code
- JWH
- PMA / PMN Number
- K040770
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). CONCOMITANT DEVICES : BIOMET CC CRUCIATE TRAY 79MM, CATALOG: #141235, LOT: #475040, VNGD PS OPEN INTL FEM RT 75, CATALOG: #183114, LOT: #488570. THE RESULTS OF THE INVESTIGATION ARE AS FOLLOWS: NO PRODUCT WAS RETURNED; VISUAL AND DIMENSIONAL EVALUATIONS COULD NOT BE PERFORMED. PHOTOGRAPHS PROVIDED CONFIRMS PATELLA PEG IS FRACTURED OFF THE DEVICE. REVIEW OF DEVICE HISTORY RECORDS IDENTIFIED NO DEVIATIONS OR ANOMALIES DURING MANUFACTURING. THIS DEVICE IS USED FOR TREATMENT. NO COMPATIBILITY ISSUES WERE NOTED. NO MEDICAL RECORDS WERE PROVIDED. X-RAYS WERE PROVIDED BUT NOT SENT FOR REVIEW. POLY IS NOT VISIBILE ON X-RAY AND WOULD LIKELY NOT ENHANCE INVESTIGATION. ROOT CAUSE CANNOT BE DETERMINED. NO CORRECTIVE ACTIONS, PREVENTIVE ACTIONS, OR FIELD ACTIONS RESULTED AFTER INVESTIGATION OF THIS EVENT. THIS COMPLAINT IS CONFIRMED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
IT WAS REPORTED THAT THE PS POLY POST WAS BROKEN. PATIENT STATED TURNING OVER IN BED AND HEARING A POPPING. PATIENT WAS REVISED MIDYEAR 2021. ATTEMPTS HAVE BEEN MADE, BUT NO FURTHER INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 865715 | ARCOM SERIES A PAT W/WR STD 31 1 PEG | PROSTHESIS, KNEE | JWH | ZIMMER BIOMET, INC. | N/A | 792970 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |