FDA Adverse Event Injury Summary report: N

VERCISE GEVIA

MDR report key: 11735581 · Received April 27, 2021

Report

Report Number
3006630150-2021-01805
Event Type
Injury
Date Received
April 27, 2021
Date of Event
March 29, 2021
Report Date
May 27, 2021
Manufacturer
BOSTON SCIENTIFIC NEUROMODULATION
Product Code
NHL
UDI-DI
08714729984443
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

CORRECTION TO BLOCK B5.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT'S RIGHT BURR HOLE INCISION WAS OPEN AND PRODUCING DISCHARGE. IT WAS ASSESSED THAT THE PATIENT HAD AN INFECTION AND UNDERWENT AN EXPLANT PROCEDURE IN WHICH ALL DEVICES WERE EXPLANTED. THE EXPLANTED DEVICES WERE DISCARDED AND WERE NOT BE RETURNED FOR ANALYSIS. A CULTURE WAS TAKEN, BUT THE RESULTS ARE UNKNOWN. THE PATIENT WAS PLACED ON ANTIBIOTICS, AND HAS RECOVERED.

Additional Manufacturer Narrative · 1

ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENTS INVOLVED IN THE EVENT: PRODUCT FAMILY: DBS-LINEAR LEADS: UPN: M365DB2202450, MODEL: DB-2202-45, SERIAL: (B)(4), BATCH: 7072760. PRODUCT FAMILY: DBS-LINEAR LEADS: UPN: M365DB2202450, MODEL: DB-2202-45, SERIAL: (B)(4), BATCH: 7075238. PRODUCT FAMILY: DBS-LEAD FIXATION: UPN: M365DB4600C0, MODEL: DB-4600C, BATCH: 26050826. PRODUCT FAMILY: DBS-EXTENSION: UPN: M365NM3138550, MODEL: NM-3138-55, SERIAL: (B)(4), BATCH: 7083006. PRODUCT FAMILY: DBS-EXTENSION: UPN: M365NM3138550, MODEL: NM-3138-55, SERIAL: (B)(4), BATCH: 7084234.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT'S RIGHT BURR HOLE INCISION WAS OPEN AND PRODUCING DISCHARGE. IT WAS ASSESSED THAT THE PATIENT HAD AN INFECTION AND UNDERWENT AN EXPLANT PROCEDURE IN WHICH ALL DEVICES WERE EXPLANTED. A CULTURE WAS TAKEN, BUT THE RESULTS ARE UNKNOWN. THE PATIENT WAS PLACED ON ANTIBIOTICS, AND HAS RECOVERED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
626651 VERCISE GEVIA STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS NHL BOSTON SCIENTIFIC NEUROMODULATION DB-1200-S 745157 08714729984443

Patients

Seq Age Sex Outcome Treatment
1 69 YR Required Intervention