FDA Adverse Event Malfunction Summary report: N

ARTICULAR SURFACE FIXED BEARING POSTERIOR STABILIZED (PS) RIGHT 11 MM HEIGHT

MDR report key: 11651901 · Received April 12, 2021

Report

Report Number
3007963827-2021-00084
Event Type
Malfunction
Date Received
April 12, 2021
Date of Event
March 23, 2021
Report Date
May 27, 2021
Manufacturer
ZIMMER ORTHOPAEDIC MFG. LTD.
Product Code
JWH
UDI-DI
00889024241190
PMA / PMN Number
K113369
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. VISUAL INSPECTION OF THE RETURNED DEVICE EXHIBITS PITTING AND NICKS ON THE CONDYLE SURFACE AND THE DOVE TAIL IS FLARED OUT. THE DHR WAS REVIEWED AND NO DISCREPANCIES RELEVANT TO THE REPORTED EVENT WERE FOUND. MEDICAL RECORDS WERE NOT PROVIDED. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Additional Manufacturer Narrative · 1

(B)(4). CONCOMITANT MEDICAL PRODUCTS: 42532007902 - PSN TIB STM 5 DEG SZ G R - 64830896. PRODUCT HAS BEEN RECEIVED BY ZIMMER BIOMET AND THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ARTICULAR SURFACE WOULD NOT ENGAGE WITH THE TIBIAL BASE PLATE. AFTER SEVERAL ATTEMPTS, ANOTHER ARTICULAR SURFACE WAS OPENED. THE SURGICAL TECHNIQUE WAS UTILIZED. THERE WAS NO SURGICAL DELAY. THERE WAS NO IMPACT OR CONSEQUENCES TO THE PATIENT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
548769 ARTICULAR SURFACE FIXED BEARING POSTERIOR STABILIZED (PS) RIGHT 11 MM HEIGHT PROSTHESIS, KNEE JWH ZIMMER ORTHOPAEDIC MFG. LTD. N/A 64094209 00889024241190

Patients

Seq Age Sex Outcome Treatment
1