FDA Adverse Event Injury Summary report: N

STEMMED NONAUGMENTABLE TIBIAL COMPONENT OPTION CR/PS/LPS SIZE 4

MDR report key: 11528511 · Received March 19, 2021

Report

Report Number
0001822565-2021-00785
Event Type
Injury
Date Received
March 19, 2021
Date of Event
February 4, 2021
Report Date
March 19, 2021
Manufacturer
ZIMMER BIOMET, INC.
Product Code
JWH
PMA / PMN Number
K173057
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CONCOMITANT MEDICAL PRODUCT: ARTICULAR SURFACE SIZE EF 10 MM HEIGHT, ITEM# 00596203210, LOT# 61447452. NO PRODUCT WAS RETURNED OR PICTURES PROVIDED; THEREFORE, VISUAL AND DIMENSIONAL EVALUATIONS COULD NOT BE PERFORMED. RADIOGRAPHS WERE PROVIDED AND REVIEWED BY A HEALTH CARE PROFESSIONAL. REVIEW OF THE AVAILABLE RECORDS IDENTIFIED THE FOLLOWING: ZONE OF PERIPROSTHETIC LUCENCY BETWEEN THE TIBIAL STEM TIP AND CEMENT/BONE MAY INDICATE LOOSENING. DEVICE HISTORY RECORD (DHR) WAS REVIEWED FOR DEVIATIONS AND/ OR ANOMALIES WITH NO RELATED DEVIATIONS / ANOMALIES IDENTIFIED. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED. PER PACKAGE INSERT 87-6203-453-01, LOOSENING IS A KNOWN ADVERSE EFFECT OF THE SYSTEM. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT WAS REVISED APPROXIMATELY TEN YEARS AND TEN MONTHS¿ POST IMPLANTATION DUE TO TIBIAL LOOSENING. THERE IS NO ADDITIONAL INFORMATION AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
428830 STEMMED NONAUGMENTABLE TIBIAL COMPONENT OPTION CR/PS/LPS SIZE 4 PROSTHESIS, KNEE JWH ZIMMER BIOMET, INC. N/A 61390963

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R