91341 TELEMETRY TRANSMITTER, 1400 MHZ
Report
- Report Number
- 3023361-2008-00039
- Event Type
- Malfunction
- Date Received
- August 22, 2008
- Date of Event
- April 27, 2006
- Report Date
- August 22, 2008
- Manufacturer
- SPACELABS HEALTHCARE
- Product Code
- DRG
- PMA / PMN Number
- K925510
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THE CUSTOMER REPORTED AN INCIDENT IN WHICH, UPON BEING POWERED UP, A TELEMETRY TRANSMITTER SET TO TRANSMIT ON A CERTAIN FREQUENCY CHANNEL (FOR CLARITY "CHANNEL 1") WOULD BEGIN TRANSMITTING ON THE WRONG RADIO FREQUENCY CHANNEL (FOR CLARITY "CHANNEL 2"). WHEN THIS OCCURS, IF ANOTHER TELEMETRY RECEIVER AT THE SITE HAS BEEN TUNED TO RECEIVE TRANSMISSIONS ON THIS OTHER CHANNEL - CHANEL 2, THE PATIENT'S WAVEFORM WOULD BE DISPLAYED ON THE CENTRAL MONITOR AS HAVING BEEN TRANSMITTED ON CHANNEL 2. PRIOR TO RECEIPT OF THIS COMPLAINT, SPACELABS HAVE INITIATED A RECALL TO ADDRESS PREVIOUSLY REPORTED INCIDENTS OF CHANNEL SWITCHING. AT THE TIME OF THIS COMPLAINT, THE CUSTOMER'S PRODUCTS HAD NOT YET BEEN UPDATED IN CONNECTION WITH THE RECALL CORRECTIVE ACTION. THIS RECALL IS STILL IN PROCESS. NOTE: THIS REPORT IS BEING SUBMITTED AS A RESULT OF A REEVALUATION OF PAST COMPLAINTS BY SPACELABS. ALTHOUGH A REPORT HAD NOT PREVIOUSLY BEEN FILED FOR THIS INCIDENT, WE HAVE CONCLUDED THAT A REPORT IS APPROPRIATE. WE ARE NOW REPORTING THIS INCIDENT IN ACCORDANCE WITH 21 CFR 803. SPACELABS HAD PREVIOUSLY REPORTED THIS CHANNEL SWITCHING ISSUE TO THE FDA PER 21 CFR 806 AS PART OF A CORRECTIVE ACTION.
THE CUSTOMER COMPLAINED ABOUT A HIGH FAILURE RATE OF 1400 MHZ TRANSMITTERS DURING THE PRIOR YEAR. INVESTIGATION REVEALED THAT THREE OF THE RETURNED TRANSMITTERS EXHIBITED CHANNEL SWITCHING, FOR WHICH SPACELABS HAS AN OPEN FCA IN PROGRESS. THIS REPORT IS FOR ONE OF THE HOSPITAL'S THREE TRANSMITTERS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 91341 TELEMETRY TRANSMITTER, 1400 MHZ | NONE | DRG | SPACELABS HEALTHCARE | 91341 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |