FDA Adverse Event Injury Summary report: N

SIGMA STAB GVF INS 2.5 15MM

MDR report key: 1063273 · Received June 18, 2008

Report

Report Number
1818910-2008-02425
Event Type
Injury
Date Received
June 18, 2008
Date of Event
May 23, 2008
Report Date
May 23, 2008
Manufacturer
DEPUY-ORTHOPAEDICS, INC., DEPUY-RAYNHAM
Product Code
JWH
PMA / PMN Number
K033272
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EXAMINATION OF THE RETURNED PRODUCT/PACKAGING AND REVIEW OF THE DEVICE HISTORY RECORDS CONFIRMED THAT THE SIZE 3 PRODUCT WAS LABELED AS SIZE 2. THE ROOT CAUSE IS ATTRIBUTED TO HUMAN ERROR IN THE LABEL PROCESS. DEPUY CONSIDERS THIS INVESTIGATION CLOSED.

Description of Event or Problem · 1

THE PT WAS REVISED BECAUSE OF THIGH PAIN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SIGMA STAB GVF INS 2.5 15MM 87JWH JWH DEPUY-ORTHOPAEDICS, INC., DEPUY-RAYNHAM NA Z55LC4000

Patients

Seq Age Sex Outcome Treatment
1 62 YR Required Intervention