FDA Adverse Event
Malfunction
Summary report: N
GENICON SUCTION IRRIGATION SYSTEM
MDR report key: 1040632
·
Received May 7, 2008
Report
- Report Number
- 1064003-2008-08002
- Event Type
- Malfunction
- Date Received
- May 7, 2008
- Date of Event
- April 22, 2008
- Report Date
- May 7, 2008
- Manufacturer
- GENICON
- Product Code
- GCX
- PMA / PMN Number
- K041967
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
FIELD INSPECTION: THE HEALTH CARE PROVIDER IS REQUESTED TO GENTLY TUG ON THE SUCTION AND IRRIGATION VALVE CAPS PRIOR TO USE OF THE LOT NUMBERS NOTED ABOVE. IF THE DEVICE IS DEFECTIVE, THE CAP(S) WILL EASILY PULL FREE FROM THE HANDPIECE. ALL DEFECTIVE PRODUCT IS TO BE RETURNED TO THE MFR.
Description of Event or Problem · 1
THE CAPS ON THE TRUMPET VALVES OF THE SUCTION IRRIGATION DEVICES COULD FALL OFF, RESULTING IN THE MALFUNCTION OF THE SUCTION IRRIGATION DEVICE. THE DEVICE WILL NOT PERFORM AS INTENDED IF THE VALVE CAPS ARE REMOVED. NO PT INJURIES OR DEATHS HAVE OCCURRED. THE DEVICE WILL NOT PERFORM AS INTENDED, AND THE DEFECT IS EVIDENT WITHIN THE PACKAGE OR AS SOON AS IT IS TAKEN OUT OF THE PACKAGE, PRIOR TO USE IN THE SURGERY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GENICON SUCTION IRRIGATION SYSTEM | SUCTION IRRIGATION VALVE ASSEMBLY | GCX | GENICON | 441071225 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |