NEUROVENT-P-TEMP
Report
- Report Number
- 3006942548-2020-00004
- Event Type
- Injury
- Date Received
- August 10, 2020
- Date of Event
- July 8, 2020
- Report Date
- July 13, 2020
- Manufacturer
- RAUMEDIC AG
- Product Code
- GWM
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- RISK MANAGER
Narratives
MANUFACTURER STATEMENT: FOR EVALUATION OF THE MALFUNCTION DHR DOCUMENTS WERE REVIEWED. THEY DEMONSTRATE THAT THE CATHETER NEUROVENT-P-TEMP (SN (B)(4)) MET SPECIFICATION DURING MANUFACTURING. FINAL INSPECTION OF THE FINISHED CATHETER WAS PASSED. THIS DEMONSTRATES THAT THE CATHETER HAS BEEN MANUFACTURED AND SOLD IN CONFORMANCE TO RELEVANT SPECIFICATIONS. FURTHERMORE THE RETURNED CATHETER WAS INVESTIGATED. THE DESCRIBED BEHAVIOR OF SHOWING NEGATIVE ICP PRESSURE VALUES COULD RECREATED. THE RESULTS OF THE PRESSURE SENSITIVITY TEST DEMONSTRATE THAT THE CATHETER IS STILL MEASURING IN A PROPER WAY, DESPITE THE OFFSET. THE VISUAL TEST OF THE MEASUREMENT WINDOW HAS SHOWN, THAT THE SILICONE LAYER OVER THE MICROCHIP WAS CROOKED. THIS EXPLAINS THE DESCRIBED OFFSET. THIS DEFORMATION COULD BE CAUSED BY A HIGH MECHANICAL FORCE APPLIED ON THE SILICONE COVER OVER THE MICROCHIP. TO EXCLUDE ANY FURTHER POSSIBLE REASON FOR THIS OFFSET, IT WAS INNSIGATED IF ANY MECHANICAL FORCES WERE APPLIED ON THE CABLE LUMEN. THE CATHETER DID'NT SUFFER ANY, AS CLARIFIED IN THE TIGHTNESS TEST. THE RESULTS OF THE ANALYSIS INDICATE A USER ERROR, SINCE PRECAUTIONARY MEASURES AND NOTES ACC. SECTION 6 OF THE INSTRUCTIONS FOR USE ZWO-013 HAVE NOT BEEN SUFFICIENTLY OBSERVED. PRECAUTION: THE PRECISION PRESSURE CATHETERS ARE PRECISION MEASURING INSTRUMENTS AND SHOULD BE TREATED ACCORDINGLY. PRECAUTION: PRIOR AND DURING THE EXAMINATION, THE PRESSURE CATHETER MUST NOT BE BENT, STRETCHED OR SQUEEZED AS WELL AS MANIPULATED BY SURGICAL TOOLS.
AFTER IMPLANTATION OF THE CATHETER NEUROVENT-P-TEMP ON (B)(6) 2020 THE ICP SIGNAL SHOWED INITIALLY UNDERSTANDABLE ICP VALUES AND A GOOD CURVE. AFTER 12 HOURS THE VALUES WERE INTERMITTENTLY AND AFTER 26 HOURS ONLY NEGATIVE. THE ICP CURVE WERE STILL WELL BUT BELOW ZERO. DESPITE ZERO ADJUSTMENT AND LOOSENING THE BOLT NO CHANGE. ON (B)(6) 2020 THE CLINIC DECIDED TO REPLACE THE CATHETER USING THE SAME BOREHOLE (SAME BOLT). THE NEW CATHETER WORKED FINE, THE VALUES RETURNED TO NORMAL. ACCORDING TO THE CLINIC THE HEALTH CONDITION OF THE PATIENT WAS NOT INFLUENCED BY THIS. THE CLINIC DECIDED AN ADDITIONAL SURGERY WAS NECESSARY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 848372 | NEUROVENT-P-TEMP | NEUROLOGICAL CATHETER | GWM | RAUMEDIC AG | 094268-001 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 38 YR | Required Intervention |