FDA Adverse Event Injury Summary report: N

AMICUS APHERSIS KIT WITH DOUBLE NEEDLE W/INL

MDR report key: 1004836 · Received February 29, 2008

Report

Report Number
2648979-2008-00002
Event Type
Injury
Date Received
February 29, 2008
Date of Event
January 9, 2008
Report Date
February 29, 2008
Manufacturer
FENWAL, INC.
Product Code
GKT
PMA / PMN Number
BK050038
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE SAMPLE WAS NOT RETURNED TO FENWAL FOR EVAL. ADDITIONAL INVESTIGATION ACTIVITIES: A TWO YR REVIEW OF PROD CODE 4R2341 DID NOT IDENTIFY ANY SIMILAR REPORTS. A REVIEW OF THE COMPLAINT LOT N07I20010, DID NOT REVEAL ANY SIMILAR REPORTS OF SUSPECTED BACTERIAL CONTAMINATION. THE MFG FACILITIES REVIEWED BATCH RECORDS FOR THIS AND SURROUNDING BATCHES, WHICH INCLUDED STERILITY TEST RESULTS, PYROGEN TESTING, BIOBURDEN, INPROCESS & FINAL CHEMICAL ANALYSIS, PARTICLE ANALYSIS, SOLUTION FILTER TESTING, WITH NO EXCEPTIONS NOTED AND ALL RESULTS WITHIN APPLICABLE SPEC. PER MEDICAL DIR; SEPTIC REACTIONS ARE RECOGNIZED ADVERSE EVENTS ASSOCIATED WITH PLATELETS TRANSFUSION. THE USUALLY OCCUR DUE TO BACTERIAL CONTAMINATION DURING COLLECTION. RECENT DATA INDICATE THAT BACTERIAL CULTURES ON DAY ONE AFTER COLLECTION MISSES ABOUT 50% OF BACTERIAL CONTAMINATION UNITS. BASED ON INFO REC'D AND REVIEWED NO CASUALITY RELATIONSHIP BETWEEN THIS EVENT AND THE FENWAL AMICUS APHERESIS KIT HAS BEEN IDENTIFIED. THIS CONSTITUTES A FINAL REPORT.

Description of Event or Problem · 1

IN 2008, A REPORT WAS REC'D THAT A SINGLE PLATELET DONATION OF 600 ML, WAS SPLIT INTO 3 PRODS AND TRANSFUSED TO 3 SEPARATE PTS. EACH PT EXPERIENCED SYMPTOMS FOLLOWING THEIR TRANSFUSIONS THAT THE CUSTOMER FEELS ARE ASSOCIATED WITH THE SPLIT PROD. TWO OF THE THREE PRODS WERE TESTED BY THE TRANSFUSING FACILITY AND TESTED POSITIVE FOR COAG-NEG STAPH. THIS REPORT IS FOR PT # 2. PP2: ADULT MALE WITH HEMATOLOGIC MALIGNANCY AND LOW BLOOD COUNTS FROM CHEMOTHERAPY EXPERIENCED AN APPARENT FEBRILE REACTION TO PLATELETS AND STARTED ANTIBIOTICS. SYMPTOMS RESOLVED AND HIS CHEMOTHERAPY REGIMEN IS STATE TO BE CONTINUING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AMICUS APHERSIS KIT WITH DOUBLE NEEDLE W/INL AMICUS DISPOSABLE KIT GKT FENWAL, INC. NA N07I20010

Patients

Seq Age Sex Outcome Treatment
1 YR Required Intervention AMICUS SEPARATOR