AMICUS APHERSIS KIT WITH DOUBLE NEEDLE W/INL
Report
- Report Number
- 2648979-2008-00002
- Event Type
- Injury
- Date Received
- February 29, 2008
- Date of Event
- January 9, 2008
- Report Date
- February 29, 2008
- Manufacturer
- FENWAL, INC.
- Product Code
- GKT
- PMA / PMN Number
- BK050038
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
THE SAMPLE WAS NOT RETURNED TO FENWAL FOR EVAL. ADDITIONAL INVESTIGATION ACTIVITIES: A TWO YR REVIEW OF PROD CODE 4R2341 DID NOT IDENTIFY ANY SIMILAR REPORTS. A REVIEW OF THE COMPLAINT LOT N07I20010, DID NOT REVEAL ANY SIMILAR REPORTS OF SUSPECTED BACTERIAL CONTAMINATION. THE MFG FACILITIES REVIEWED BATCH RECORDS FOR THIS AND SURROUNDING BATCHES, WHICH INCLUDED STERILITY TEST RESULTS, PYROGEN TESTING, BIOBURDEN, INPROCESS & FINAL CHEMICAL ANALYSIS, PARTICLE ANALYSIS, SOLUTION FILTER TESTING, WITH NO EXCEPTIONS NOTED AND ALL RESULTS WITHIN APPLICABLE SPEC. PER MEDICAL DIR; SEPTIC REACTIONS ARE RECOGNIZED ADVERSE EVENTS ASSOCIATED WITH PLATELETS TRANSFUSION. THE USUALLY OCCUR DUE TO BACTERIAL CONTAMINATION DURING COLLECTION. RECENT DATA INDICATE THAT BACTERIAL CULTURES ON DAY ONE AFTER COLLECTION MISSES ABOUT 50% OF BACTERIAL CONTAMINATION UNITS. BASED ON INFO REC'D AND REVIEWED NO CASUALITY RELATIONSHIP BETWEEN THIS EVENT AND THE FENWAL AMICUS APHERESIS KIT HAS BEEN IDENTIFIED. THIS CONSTITUTES A FINAL REPORT.
IN 2008, A REPORT WAS REC'D THAT A SINGLE PLATELET DONATION OF 600 ML, WAS SPLIT INTO 3 PRODS AND TRANSFUSED TO 3 SEPARATE PTS. EACH PT EXPERIENCED SYMPTOMS FOLLOWING THEIR TRANSFUSIONS THAT THE CUSTOMER FEELS ARE ASSOCIATED WITH THE SPLIT PROD. TWO OF THE THREE PRODS WERE TESTED BY THE TRANSFUSING FACILITY AND TESTED POSITIVE FOR COAG-NEG STAPH. THIS REPORT IS FOR PT # 2. PP2: ADULT MALE WITH HEMATOLOGIC MALIGNANCY AND LOW BLOOD COUNTS FROM CHEMOTHERAPY EXPERIENCED AN APPARENT FEBRILE REACTION TO PLATELETS AND STARTED ANTIBIOTICS. SYMPTOMS RESOLVED AND HIS CHEMOTHERAPY REGIMEN IS STATE TO BE CONTINUING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AMICUS APHERSIS KIT WITH DOUBLE NEEDLE W/INL | AMICUS DISPOSABLE KIT | GKT | FENWAL, INC. | NA | N07I20010 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | YR | Required Intervention | AMICUS SEPARATOR |