The Fresenius APD Luer-Lock Adapter, Catalog number 050-95018, is designed to connect a peritoneal dialysis bag equipped with a luer-lock connector to a Fresenius Automated Peritoneal Dialysis (APD) cycler set that uses the Fresenius Safe-Lock¿ connector (e.g., Liberty¿ cycler, Newton IQO cycler).
FDA Enforcement
Class II
·Terminated·Fresenius Medical Care Holdings, Inc.·July 28, 2021
Hemosphere System with the following three components: HEM1 Advanced Monitor, HEMSGM10 (Swan-Ganz Module), and HEMOXSC100 (Oximetry Smart Cable). Product Usage: The HemoSphere Advanced Monitor (HEM1) is intended to be used in combination with a compatible Edwards hemodynamic monitoring technology module and/or cable and accompanying Edwards accessories and/or disposables. It provides Edwards hemodynamic monitoring technologies through modular connections. The HemoSphere Swan-Ganz Module (HEMSGM10) is intended to be used by qualified personnel or trained clinicians in a critical care environment in a hospital setting. It is a multi-use device that works in conjunction with the HemoSphere Advanced Monitor, patient CCO cable and compatible Swan-Ganz catheter to acquire and process temperature data supporting intermittent cardiac output (iCO), continuous cardiac output (CCO) and right ventricular end diastolic volume (EDV) monitoring. The HemoSphere Oximetry Smart Cable (HEMOXSC100) is intended to be used by qualified personnel or trained clinicians in a critical care environment in a hospital setting. It is a multi-use device that connects with a compatible Edwards monitor on one end and any approved Edwards oximetry catheter on the other end to continuously measure venous oxygen saturation by reflectance spectrophotometry. LEDs within the oximetry cable transmit light fiber optically to the distal end of the catheter. The amount of light absorbed, refracted, and reflected depends on the relative amounts of oxygenated and deoxygenated hemoglobin in the blood. This optical intensity data is gathered by the oximetry catheter, processed by the HemoSphere Oximetry Cable, and displayed on a compatible monitoring platform. Parameter output is mixed venous oxygen saturation (SvO2) or central venous oxygen saturation (ScvO2).
FDA Enforcement
Class II
·Terminated·Edwards Lifesciences, LLC·February 7, 2018
Ikaria, INOmax DSIR (Delivery System), Model 10007. Nitric oxide delivery system for use with ventilators.
FDA Enforcement
Class II
·Terminated·INO Therapeutics (dba Ikaria)·March 11, 2015
Mallinckrodt Pharmaceuticals INOmax DSIR Plus delivery system. Model 10007. Delivers INOMAX (nitric oxide for inhalation). The INOmax DSIR uses a dual-channel design to ensure the safe delivery of INOMAX. The first channel has the delivery CPU, the flow controller and the injector module to ensure the accurate delivery of NO. The second channel is the monitoring system, which includes a separate monitor CPU, the gas cells (NO, NO2, and 02 cells) and the user interface including the display and alarms. The dual-channel approach to delivery and monitoring permits INOMAX delivery independent of monitoring but also allows the monitoring system to shutdown INOMAX delivery if it detects a fault in the delivery system such that the NO concentration could become greater than 100 ppm. The delivery system can also shut down delivery if it detects certain serious problems with the monitoring system.
FDA Enforcement
Class II
·Terminated·INO Therapeutics (dba Ikaria)·February 1, 2017
INOmax DSIR, model 10007, Nitric Oxide Delivery System, with Software version 3.0.0. The INOmax DS delivery system delivers INOMAX (nitric oxide for inhalation) therapy gas and provides continuous monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.
FDA Enforcement
Class II
·Terminated·INO Therapeutics (dba Ikaria)·May 14, 2014
INOMAX DSIR Nitric Oxide delivery system, Model 10007, software version 3.1.0., delivers INOMAX (nitric oxide for inhalation) therapy gas and provides continuous monitoring of inspired O2, NO2, and NO, and comprehensive alarm systems. The primary clinical settings are Neonatal Intensive Care Units (NICU) and to transport of neonates.
FDA Enforcement
Class II
·Terminated·INO Therapeutics (dba Ikaria)·December 17, 2014
Mallinckrodt Pharmaceuticals INOflo DS, NO Gas Control System. Product Usage; Device Classification: The INOmeter is an accessory to the nitric oxide delivery device which is used for the delivery of inhaled nitric oxide to patient. An INOmeter is affixed to each cylinder of drug product. Functionality of the INOmeter: The INOmeter is affixed on top of each INOflo cylinder valve in Japan. When the healthcare provider opens the INOmeter the compressed gas cylinder valve opens to begin flow of the nitric oxide in nitrogen compressed gas. The nitric oxide gas is administered to a patient via a nitric oxide delivery device. The amount of time that gas flow administered to a patient is captured in the INOmeter. Each INOmeter is programmed to include the drug product expiration date as listed on the INOflo drug product label. The INOmeter electronically communicates to the nitric oxide delivery device via infrared technology by way of the iButton and in so communicates the expiration date of the corresponding cylinder of INOflo drug product to the delivery device. The delivery device evaluates and compares the expiration date to the current date of therapy administration. If the drug product expiration date is prior to the current date, an error and alarm is registered by the nitric oxide delivery device and administration of the expired drug to the patient is stopped after 120 seconds. If this occurs, the healthcare provider can operate the nitric oxide delivery device manually and deliver the INOflo as indicated in the Instructions For Use (IFU). The INOflo delivery device has the functionality to administer INOflo in manual mode.
FDA Enforcement
Class II
·Terminated·INO therapeutics LLC/dba ikaria·August 15, 2018
GE Healthcare Optima CT580/Discovery CT590RT scanners running software version 13BW16.3. The Discovery CT 590 RT and Optima CT 580 series of Computed Tomography systems are intended to produce cross-sectional images of the body by computer reconstruction of XRay transmission data taken at different angles and planes, including Axial, Cine, Helical (Volumetric), and Gated (Respiratory and Cardiac) acquisitions in patients of all ages.