FDA UDI In Commercial Distribution 🇺🇸 United States

SACCADOMETER

DI: 05906874392010 · Model: PLUS · OBER CONSULTING SP Z O O
Product Codes
1
GMDN Terms
1
Identifiers
1
Pkg Device Count
1

Basic Information

Brand Name
SACCADOMETER
Primary DI
05906874392010
Version / Model
PLUS
Company Name
OBER CONSULTING SP Z O O
Labeler DUNS
522393420
Distribution Status
In Commercial Distribution
Device Count in Pkg
1
Record Status
Published
Publish Date
2017-01-26
Public Version
3
Public Version Date
2018-07-06
Public Version Status
Update
Public Device Record Key
6608604e-774c-49ac-8c9a-4fa7d55887ec

Device Description

The Saccadometer Plus system measures eye movements in horizontal axis with temporal 1kHz and spatial resolution < 5 arc min. The system incorporates a built in visual target presentation using miniature laser projectors mounted on the sensor forehead plate. Saccadometer Plus is equipped with three red laser projectors with 10º separation, and 6 experiments are included in embedded software. The information about the eye movement is stored in a small, self-contained control unit, and can be downloaded for computer analysis at a later stage. Information about eye movement is acquired by analyzing the infrared light reflected from eye orbit. The left and right eye orbit are illuminated by two infrared LEDs, providing eye surface irradiance below 1mW/cm 2 (operating condition). Due to the conjugacy and synchronicity of the saccadic eye movements, the rotation of left and right eyes, can be added and averaged. The inner canthi of the left and right eyes are illuminated with the low intensity IR. The difference between the amounts of IR reflected back from the eye surfaces toward the photodetecting circuitry, carries the information about the eye position changes. Two eye landmarks are taking part in generation of the eye movement signal: the corneal bulge and the limbus (the border between darker iris and white sclera). Main contributor to the eye movement signal is the corneal bulge. Being the relative eye position measurement system, it requires to define the initial eye position. It means that at the beginning of every measurement, a steady eye fixation point needs to be provided. In Saccadometer the central fixation target serves this purpose.

Device Characteristics

Single Use
No
Prescription Use (Rx)
Yes
Over the Counter (OTC)
No
Kit
No
Combination Product
No
HCT/P
No
Contains NRL
No
Not Made with NRL
No
MRI Safety
Labeling does not contain MRI Safety Information
Direct Marking Exempt
No
PM Exempt
No
Has Serial Number
Yes
Has Lot/Batch Number
No
Has Manufacturing Date
Yes
Has Expiration Date
No
Has Donation ID
No

Sterilization

Is Sterile
No
Sterilization Prior Use
No

Product Codes

Code Name
HLL Monitor, Eye Movement

GMDN Terms

Code Name
12609 Eye movement analysis system

Identifiers

Type ID
Primary 05906874392010

Customer Contacts

Phone
+48505940787

Premarket Submissions

Submission Number Supplement Number
K152890 000

Storage Conditions

Type
Handling Environment Temperature
Temperature Range
50 – 104 Degrees Fahrenheit
Type
Handling Environment Temperature
Temperature Range
10 – 40 Degrees Celsius
Type
Storage Environment Temperature
Temperature Range
32 – 113 Degrees Fahrenheit
Type
Storage Environment Temperature
Temperature Range
0 – 45 Degrees Celsius