FDA UDI In Commercial Distribution 🇺🇸 United States

PROFILER 2

DI: 05060608880140 · Model: 1174300Z · SUN NUCLEAR CORP.
Product Codes
1
GMDN Terms
1
Identifiers
1
Pkg Device Count
1

Basic Information

Brand Name
PROFILER 2
Primary DI
05060608880140
Version / Model
1174300Z
Catalog Number
11740000Z
Company Name
SUN NUCLEAR CORP.
Labeler DUNS
122240021
Distribution Status
In Commercial Distribution
Device Count in Pkg
1
Record Status
Published
Publish Date
2019-05-09
Public Version
1
Public Version Date
2019-05-17
Public Version Status
New
Public Device Record Key
1ba9e6f3-f5c9-4433-8592-532c5411390c

Device Description

Radiation therapy QA device.

Device Characteristics

Single Use
No
Prescription Use (Rx)
Yes
Over the Counter (OTC)
No
Kit
No
Combination Product
No
HCT/P
No
Contains NRL
No
Not Made with NRL
No
MRI Safety
Labeling does not contain MRI Safety Information
Direct Marking Exempt
No
PM Exempt
No
Has Serial Number
Yes
Has Lot/Batch Number
Yes
Has Manufacturing Date
Yes
Has Expiration Date
No
Has Donation ID
No

Sterilization

Is Sterile
No
Sterilization Prior Use
No

Product Codes

Code Name
IYE Accelerator, Linear, Medical

GMDN Terms

Code Name
38399 Accelerator system quality assurance device

Identifiers

Type ID
Primary 05060608880140

Premarket Submissions

Submission Number Supplement Number
K063021 000

Storage Conditions

Type
Storage Environment Temperature
Temperature Range
-30 – 50 Degrees Celsius
Type
Storage Environment Humidity
Temperature Range
10 – 90 Percent (%) Relative Humidity
Type
Special Storage Condition, Specify
Special Conditions
Store the PROFILER 2 in an indoor, protected environment where it will not be irradiated by the direct beam and where scatter radiation will be kept to a minimum.
Type
Special Storage Condition, Specify
Special Conditions
Never expose the electronics portion of the instrument to the direct beam.
Type
Special Storage Condition, Specify
Special Conditions
Never leave the instrument on the couch during warm-up.
Type
Special Storage Condition, Specify
Special Conditions
Never store the instrument where it can accidentally be irradiated, such as a shelf in the treatment room that is irradiated by the beam when the linac is rotated.
Type
Special Storage Condition, Specify
Special Conditions
Avoid frequent use of thick buildup, which may cause scattered radiation that will damage the electronics.
Type
Special Storage Condition, Specify
Special Conditions
Keep the electronics out of the beam during concatenation, rotation, and offset.