Shockwave IVL System with the Shockwave Javelin Peripheral IVL Catheter
Basic Information
- Brand Name
- Shockwave IVL System with the Shockwave Javelin Peripheral IVL Catheter
- Primary DI
- 00195451000799
- Version / Model
- JVNPIVLF
- Catalog Number
- JVNPIVLF
- Company Name
- Shockwave Medical, Inc.
- Labeler DUNS
- 055514478
- Distribution Status
- In Commercial Distribution
- Device Count in Pkg
- 1
- Record Status
- Published
- Publish Date
- 2024-10-09
- Public Version
- 4
- Public Version Date
- 2026-01-21
- Public Version Status
- Update
- Public Device Record Key
- 8d886d72-7a6d-491e-926a-95c4cd4a8b93
Device Description
The IVL Catheter is a proprietary lithotripsy device delivered through the peripheral arterial system of the lower extremities to the site of an otherwise difficult to treat calcified stenosis. Intravascular Lithotripsy (IVL) is an interventional procedure that utilizes a fluid-filled catheter connected to power sources that generate acoustic shock waves; the shock waves modify calcified plaque in peripheral arteries. Energizing the intravascular lithotripsy device will generate acoustic pressure pulses within the target treatment site, disrupting calcium within the lesion. The Shockwave Intravascular Lithotripsy (IVL) System with the Shockwave Javelin Peripheral Intravascular Lithotripsy (IVL) Catheter consists of three main components: the IVL Catheter, the IVL Generator, and the IVL Connector Cable. The Shockwave Javelin IVL Catheter is comprised of a catheter with an integrated emitter located near the distal end to enable the localized delivery of acoustic pressure pulses in the peripheral vasculature. The emitter is radiopaque to facilitate catheter visibility under fluoroscopy and is surrounded by a fluid-filled space (IVL window) that allows for the transmission of acoustic pressure pulses. The Shockwave Javelin Peripheral IVL Catheter shaft contains a lumen to pressurize and a lumen to de-pressurize the catheter with saline (the medium to create IVL), a guidewire lumen, and a lithotripsy emitter. The emitter is enclosed within a non-expandable polymer fluid-filled space (i.e., IVL window) containing saline that is connected to the proximal shaft, inlet and outlet ports, and is tapered down to the distal tip of the catheter. The IVL window is located near the distal tip of the catheter. The emitter is radiopaque to facilitate catheter visibility under fluoroscopy and is surrounded by the IVL window that allows for the transmission of acoustic pressure pulses. The IVL window is designed to provide a static catheter profile.
Device Characteristics
- Single Use
- Yes
- Prescription Use (Rx)
- Yes
- Over the Counter (OTC)
- No
- Kit
- No
- Combination Product
- No
- HCT/P
- No
- Contains NRL
- No
- Not Made with NRL
- No
- MRI Safety
- Labeling does not contain MRI Safety Information
- Direct Marking Exempt
- No
- PM Exempt
- No
- Has Serial Number
- No
- Has Lot/Batch Number
- Yes
- Has Manufacturing Date
- Yes
- Has Expiration Date
- Yes
- Has Donation ID
- No
Sterilization
- Is Sterile
- Yes
- Sterilization Prior Use
- No
Product Codes
| Code | Name | Medical Specialty | Regulation # | Device Class |
|---|---|---|---|---|
| PPN | Percutaneous Catheter, Ultrasound | Cardiovascular | 870.1250 | 2 |
| OEZ | Cardiovascular Procedure Kit | Cardiovascular | 870.1200 | 2 |
| JAA | System, X-Ray, Fluoroscopic, Image-Intensified | Radiology | 892.1650 | 2 |
GMDN Terms
| Code | Name | Definition | Implantable | Status |
|---|---|---|---|---|
| 12535 | Medical equipment/instrument drape, single-use | A flexible polymer sheet/bag/sleeve designed to form a protective water-resistant enclosure around a piece of medical equipment/instrument (e.g., C-arm, surgical laser/camera/microscope, examination table/chair) and/or components (e.g., cables, tubes, headrest). It is intended as a hygienic barrier to protect the equipment/instrument from soiling and/or contamination when used inside, or within the vicinity of, a hygienic area; it may have patient contact, however it is not intended to be introduced into a body cavity (noninvasive). It is neither a dedicated handle cover, device cap nor an ultrasound transducer cover. This is a single-use device. | No | Active |
| 67698 | Intravascular lithotripsy system catheter, non-balloon | An electrically-powered, flexible tube, which does not incorporate a balloon, intended to be introduced into a peripheral artery (excludes coronary and cerebral arteries) to emit ultrasonic acoustic pressure waves at the distal tip for the modification and fragmentation of calcified plaque, to improve vessel compliance. It is used in conjunction with a dedicated control unit/generator (not included), and includes transducers intended to convert electrical to ultrasonic energy. This is a single-use device. | No | Active |
Identifiers
| Type | ID | Issuing Agency | Package Type | Qty per Pkg | Pkg Status | Pkg Discontinue Date |
|---|---|---|---|---|---|---|
| Primary | 00195451000799 | GS1 |
Customer Contacts
- Phone
- 1-877-775-4846
- [email protected]
Premarket Submissions
| Submission Number | Supplement Number |
|---|---|
| K242213 | 000 |
Device Sizes
| Type | Value | Unit | Text |
|---|---|---|---|
| Outer Diameter | 1.5 | Millimeter | |
| Catheter Working Length | 150 | Centimeter | |
| Introducer Sheath Compatibility | 5 | French | |
| Guidewire Diameter | 0.014 | Inch |
Storage Conditions
- Type
- Special Storage Condition, Specify
- Special Conditions
- Keep away from sunlight, keep dry