FDA Recall
Terminated
Sonopet Ultrasonic Surgical System, Model UST-2001, Manufacturer Miwatec Co. LTD, Tokyo, Japan.
Recall: z-2640-2010
·
Initiated July 15, 2010
Recall
- Recall Number
- z-2640-2010
- Event Number
- 56652
- Firm
- Stryker Instruments Div. of Stryker Corporation
- FEI Number
- 1811755
- Product Code
- LFL
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- July 15, 2010
- Posted
- October 1, 2010
- Terminated
- October 19, 2011
- Address
- 4100 E. Milham Ave., Kalamazoo, MI, 49001
Description
Sonopet Ultrasonic Surgical System, Model UST-2001, Manufacturer Miwatec Co. LTD, Tokyo, Japan.
Reason
Stryker Instruments acquired the Sonopet Ultrasonic Surgical System from Mutoh co. Ltd on 3/29/2010. Stryker determined that the Sonopet consode did not meet the IEC-60601-1 standard. Stryker decided to recall the Sonopet consode.
Action
The firm to send URGENT MEDICAL DEVICE RECALL NOTIFICATIONS to customers dated 7/15/2010. The customers were instructed to continue to use the device. Stryker technicians were dispatched to replace a board with a new board.
Distribution
Wordwide
Quantity
1161