FDA Recall Open, Classified

Atellica IM aTPO Assay, Material Number 10995466 (100 Test) and 10995467 (500 Test)

Recall: Z-3305-2024 · Initiated August 21, 2024

Recall

Recall Number
Z-3305-2024
Event Number
95259
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2432235
Product Code
JZO
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
August 21, 2024
Posted
September 27, 2024
Address
511 Benedict Ave, Tarrytown, NY, 10591-5005

Description

Atellica IM aTPO Assay, Material Number 10995466 (100 Test) and 10995467 (500 Test)

Reason

Positive bias in patient results at and below the cut-off of 60 U/mL with the lots listed in the table below when compared to alternate lots on the Atellica IM analyzer.

Action

An Urgent Medical Device Correction notification letter dated 8/21/24 was sent to customers. Customer Actions " Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. " Perform the actions provided below: o Discontinue use of and discard the kit lots listed in the table above (Products Section). o You may request no-charge replacement product from your local Siemens Healthineers or distributor office. Please review your inventory of these products to determine your laboratory s replacement needs and to provide information to Siemens Healthineers for reporting to the authorities. o Complete and return the Field Correction Effectiveness Check and indicate product replacement needs on the form attached to this letter within 30 days. o Please retain this letter with your laboratory records and forward this letter to those who may have received this product. You may request no-charge replacement product from your local Siemens Healthineers or distributor office. Unaffected reagent lots are available for ordering. Please send a scanned copy of the completed form via email to: [email protected] Or to fax this completed form to the Customer Care Center at: (312) 275-7795 We apologize for the inconvenience this situation may cause. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.

Distribution

Worldwide distribution - US Nationwide.

Quantity

9783 kits