FDA Recall Open, Classified

Origio Sperm Wash Medium Pack of Ten 10ml bottles - Intended for washing of spermatozoa Part Number: 84051010A

Recall: Z-3296-2024 · Initiated February 16, 2024

Recall

Recall Number
Z-3296-2024
Event Number
95246
Firm
CooperSurgical, Inc.
FEI Number
1216677
Product Code
MQL
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
February 16, 2024
Posted
September 26, 2024
Address
95 Corporate Dr, Trumbull, CT, 06611-1350

Description

Origio Sperm Wash Medium Pack of Ten 10ml bottles - Intended for washing of spermatozoa Part Number: 84051010A

Reason

Low concentration of sodium pyruvate in the wash could compromise gamete development

Action

CooperSurgical issued Urgent: Voluntary Medical Device Field Safety Notice on 2/16/24 via UPS Next Day Air via third party Sedgewick. Letter states reason for recall, health risk and action to take: 1) Inspect your inventory, quarantine any affected product, and immediately stop using the product. 2) Complete the appropriate version of the attached form (Customer Acknowledgement Form or Distributor Acknowledgement Form). Once completed, return the form to CooperSurgical as indicated at the top of the form. Note: Even if you do not have any affected product in your inventory ( or have not distributed any affected product to customers), please complete and return the enclosed form so that we may document receipt of this FSN. When the completed form is received by CooperSurgical, arrangements will be made for the return of any affected product at no additional cost to you. Credit will be applied to your account for product returned under this action. CooperSurgical is committed to high-quality, safe, and effective products. This corrective action has been initiated to ensure this failure mode does not reoccur and future potential patient harms can be avoided. If you have any questions, please feel free to reach us at +1 203-601-5200 ext. 3300 or [email protected].

Distribution

Worldwide distribution - US Nationwide.

Quantity

980 units