FDA Recall
Terminated
AXS Infinity LS Long Sheath:((6F/.088in)/ x 70cm GEN-10800-70, (6F/.088in)/ x 80cm GEN-10800-80, (6F/.088in)/ x 90cm GEN-10800-90) Made in Ireland, Sterile EO. The AXS Infinity LS Long Sheath system contents are: One (1) AXS Infinity LS Long Sheath, One (1) Hemostasis valve with extension tubing, One Long Dilator.
Recall: Z-3281-2018
·
Initiated May 2, 2018
Recall
- Recall Number
- Z-3281-2018
- Event Number
- 80652
- Firm
- Stryker Neurovascular
- FEI Number
- 3008853977
- Product Code
- DQY
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- May 2, 2018
- Address
- 47900 Bayside Pkwy, Fremont, CA, 94538-6515
Description
AXS Infinity LS Long Sheath:((6F/.088in)/ x 70cm GEN-10800-70, (6F/.088in)/ x 80cm GEN-10800-80, (6F/.088in)/ x 90cm GEN-10800-90) Made in Ireland, Sterile EO. The AXS Infinity LS Long Sheath system contents are: One (1) AXS Infinity LS Long Sheath, One (1) Hemostasis valve with extension tubing, One Long Dilator.
Reason
The AXS Infinity LS Long Sheath Dilator outer diameter (OD) specification was exceeded.
Action
On 5/2/2018, "Urgent Medical Device Customer Communication" notifications were mailed via courier. Customers were informed to complete and return a Customer Acknowledgement Form. Customers with questions are encouraged to call (510) 413-2100.
Distribution
Only U.S. Consignees: CO,NJ,NY, OH,PA, and WV.
Quantity
15 units