Brilliance 64 Computed Tomography System, Model #78231, Serial Number: 9592. The Brilliance 64 is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at difference angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories.
Recall
- Recall Number
- Z-3281-2011
- Event Number
- 59383
- Firm
- Philips Medical Systems (Cleveland) Inc
- FEI Number
- 1525965
- Product Code
- JAK
- Status
- Terminated
- Root Cause
- Other
- Initiated
- June 2, 2011
- Posted
- September 27, 2011
- Terminated
- March 15, 2013
- Address
- 595 Miner Road, Cleveland, OH, 44143-2131
Description
Brilliance 64 Computed Tomography System, Model #78231, Serial Number: 9592. The Brilliance 64 is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at difference angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories.
Component defect. Philips Medical decided to conduct a recall field /correction (rework) of the mounting key locking threaded insert of the Detector Module system (DMS) bolt on one (1) Brilliance 64 CT X-ray unit. The firm initiated this action following Philip's receipt of the customer's complaint reporting that the retaining inserts were pulling out of the rotor during the removal of the UDMS
Philips sent a Field Change Order (FCO) 72800519 on June 13, 2011 to the affected customer. A Field Service Engineer visited the customer site and installed a hardware upgrade to the affected Brilliance 64 device system (Serial Number: 9592). The correction action was completed on August 9, 2011. For questions regarding this recall call 1-440-483-7000.
Nationwide Distribution including PA
1 unit