FDA Recall Terminated

The Trabecular Metal Modular Acetabular System is a modular acetabular cup system consisting of a shell with a preassembled locking ring, a liner, and optional screws. The Shells are available in a variety of sizes and are designed to be used with Trilogy Acetabular System liners. The shell implants are used to replace the hip socket during total hip arthroplasty. The device is indicated for primary or revision surgery for rehabilitating hips damaged as a result on noninflammatory degenerative joint disease (NIDJD) or its composite diagnoses of osteoarthritis, avascular necrosis,protrusio acetabuli, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis and diastrophic variant. The device is intended for either cemented or noncemented use.

Recall: Z-3279-2011 · Initiated September 1, 2011

Recall

Recall Number
Z-3279-2011
Event Number
59503
Firm
Zimmer Inc.
FEI Number
1000220733
Product Code
JDI
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
September 1, 2011
Posted
September 27, 2011
Terminated
August 29, 2013
Address
345 E Main St, Warsaw, IN, 46580-2746

Description

The Trabecular Metal Modular Acetabular System is a modular acetabular cup system consisting of a shell with a preassembled locking ring, a liner, and optional screws. The Shells are available in a variety of sizes and are designed to be used with Trilogy Acetabular System liners. The shell implants are used to replace the hip socket during total hip arthroplasty. The device is indicated for primary or revision surgery for rehabilitating hips damaged as a result on noninflammatory degenerative joint disease (NIDJD) or its composite diagnoses of osteoarthritis, avascular necrosis,protrusio acetabuli, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis and diastrophic variant. The device is intended for either cemented or noncemented use.

Reason

Lot # 61132698 and Lot 001122467 (added 9/ 18/2012) may have a manufacturing non conformance that could prevent the devices from functioning as intended. Specifically, devices in the affected lot may have a locking ring groove depth that prevents the corresponding liner from snapping into the shell. The groove provides the locking ring room to displace when inserting a liner into the shell. Th

Action

Zimmer, Inc. sent an "URGENT MEDICAL DEVICE RECALL" letter dated July 27, 2011 to all affected customers of lot 61132698 and 9/18/2012 to only affected customer of lot 001122467. The letters describes the product, problem, and the actions to be taken by the customers. The letter instructs customers to stop using and to quarantine the affected product. An Inventory Return Certification Form was included with the letter for customers to complete and return to Zimmer, Inc. via fax at (574) 372-4265. Replacement product will be provided for all returned product. Contact Zimmer, Inc. at 1-800-613-6131 for questions regarding this recall.

Distribution

Worldwide Distribution-USA (nationwide) including the states of CA, NC,NY, PA, WA, and WI and countries of Australia, Germany, The Netherlands, and Switzerland.

Quantity

24 devices lot 61132689; 1 device lot 001122467