VASOVIEW 6 Endoscopic Vessel Harvesting System Manufactured by Maquet Cardiovascular, LLC; 170 Baytech Drive, San Jose, CA 95134. Product Usage: Endoscopic vessel harvesting systems are indicated for use in minimally invasive surgery allowing access for vessel harvesting and are primarily indicated for patients undergoing endoscopic surgery for arterial bypass.
Recall
- Recall Number
- Z-3270-2011
- Event Number
- 59865
- Firm
- Maquet Cardiovascular, LLC
- FEI Number
- 2242352
- Product Code
- GEI
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- December 17, 2009
- Posted
- September 20, 2011
- Terminated
- September 21, 2011
- Address
- 45 Barbour Pond Dr, Wayne, NJ, 07470-2094
Description
VASOVIEW 6 Endoscopic Vessel Harvesting System Manufactured by Maquet Cardiovascular, LLC; 170 Baytech Drive, San Jose, CA 95134. Product Usage: Endoscopic vessel harvesting systems are indicated for use in minimally invasive surgery allowing access for vessel harvesting and are primarily indicated for patients undergoing endoscopic surgery for arterial bypass.
During routine production quality monitoring process, it was found that UV curing adhesive was not used in a limited number of devices.
MAQUET Cardiovascular, LLC, sent an "Urgent Device Removal/Immediate Action Required" letter dated December 17, 2009, to all affected customers. The letter identified the recalled products, the problem, and the actions needed to be taken. The letters asked customers to examine their stocks (inventory) immediately to determine if they had any affected lot numbers and to discontinue dispensing (distributing) the lot. The letter states that customers are to complete the attached Field Action Response form regardless of whether or not the affected products are located in their inventory. Distributors are instructed to provide the same instructions to their customers supplied with the recalled products. The letter instructed all users to return all unused affected products to MAQUET. For customer's convenience, all affected products that are returned may be exchanged with other unaffected VASOVIEW Endoscopic Vessel Harvesting System by overnight delivery. Questions should be directed to MAQUET Cardiovascular sales representative or Customer Service at 1-888-880-2874.
Nationwide Distribution (USA) including the states of: AL, AZ, CA, CT, FL, KS, KY, IL, IN, MI, MO, ND, NY, NV, OH, OK, OR, PA, SC, TN, TX, UT, WA, and WI.
141 units