COULTER DxH Cell Lyse, Part number 628018(1L) and 628019 (5L). The COULTER DxH Cell Lyse is an erythrocyte lytic agent used for quantitative determination of hemoglobin, enumeration of NRBC, and counting and sizing leukocytes on UniCel DxH 800 Coulter Cellular Analysis System.

Recall: Z-3263-2011 · Initiated July 11, 2011

Recall

Recall Number
Z-3263-2011
Event Number
59726
Firm
Beckman Coulter Inc.
FEI Number
2050012
Product Code
GGK
Status
Terminated
Root Cause
Pending
Initiated
July 11, 2011
Posted
September 19, 2011
Terminated
June 4, 2012
Address
250 S Kraemer Blvd, Brea, CA, 92821-6232

Description

COULTER DxH Cell Lyse, Part number 628018(1L) and 628019 (5L). The COULTER DxH Cell Lyse is an erythrocyte lytic agent used for quantitative determination of hemoglobin, enumeration of NRBC, and counting and sizing leukocytes on UniCel DxH 800 Coulter Cellular Analysis System.

Reason

The recall was initiated because Beckman Coulter identified unexpected high particle counts in the lots of DxH Cell Lyse listed. These lots may generate high background counts; could result in a falsely elevated Total Nucleated Count (TNC) of up to 50 cells/L; or may also result in recovery of TNC values that were higher than expected for Level 1 of the Body Fluid Control.

Action

The recall communication was initiated on 08/01/2011 with Beckman Coulter forwarding a Product Corrective Action (PCA) letter with attached PCA Response Form (via US Postal Service for US customers) to all customers, who purchased Coulter DxH Cell Lyse. The letter provides the customers with an explanation of the problem identified and an action to be taken. Customers were instructed to contact Beckman Coulter for product replacement. In addition, customers were instructed to complete and return the enclosed PCA Response Form within 10 days. Customers with questions regarding this Product Corrective Action were instructed to call Beckman Coulter Customer Service at 800-526-7694 in the United States or Canada, or contact their Beckman Coulter representative.

Distribution

Worldwide Distribution -- Australia, Belgium, Brunei Darussalam, Canada, France, Germany, Hong Kong, Hungary, India, Italy, Japan, Macao, Malaysia, Netherlands, Norway, Philippines, Qatar, Russian Federation, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, and the United States.

Quantity

4913 units total (2245 in US)