FDA Recall Terminated

Model 3300 LATITUDE(TM) Programming System Product Usage - The Boston Scientific LATITUDE Programming System is intended for use in hospital and clinical environments to communicate with Boston Scientific implantable systems. The software in use controls all communication functions for the Pulse Generator (PG). Detailed software applications instructions can be found in the associated product literature for the PG being interrogated.

Recall: Z-3260-2018 · Initiated September 10, 2018

Recall

Recall Number
Z-3260-2018
Event Number
81006
Firm
Boston Scientific Corporation
FEI Number
2124215
Product Code
LWS
Status
Terminated
Root Cause
Device Design
Initiated
September 10, 2018
Terminated
September 22, 2021
Address
4100 Hamline Ave N, Bldg 3, Saint Paul, MN, 55112-5700

Description

Model 3300 LATITUDE(TM) Programming System Product Usage - The Boston Scientific LATITUDE Programming System is intended for use in hospital and clinical environments to communicate with Boston Scientific implantable systems. The software in use controls all communication functions for the Pulse Generator (PG). Detailed software applications instructions can be found in the associated product literature for the PG being interrogated.

Reason

There is a potential for the Pacing System Analyzer (PSA) to exhibit unintended cross-chamber stimulation.

Action

The firm initiated the recall on 09/10/2018 to the US sales representatives by email. The sales representatives were asked to hand deliver the notice to the hospitals when the programmer is retrieved for that location. The sales representatives will electronically confirm the hand delivery of the letter to each affected customer after it occurs.

Distribution

Worldwide distribution - US Nationwide in the State of FL, UT, and VA, and countries of Austria, Czech Republic, Germany, Greece, Netherlands, Norway, Portugal, Spain, Sweden, United Kingdom, Switzerland

Quantity

74 units